LD-560 HPLC System for HbA1c Testing
The Fully Automated HbA1c Analyzer is used in conjunction with suitable reagents for the quantitative detection of glycosylated hemoglobin
(HbA1c) in human blood.
Positioning:10-30 daily sample
Test Speed: Within 90s/T
Principle
Using high performance liquid chromatography (HPLC), the stationary
phase used is a weakly acidic cation exchanger. Glycosylation of
hemoglobin will result in the loss of cations on the surface of
hemoglobin without being charged, and non glycosylated hemoglobin
has a positive charge. According to the different charged
properties of hemoglobin, eluents with different pH values and
ionic strengths are used sequentially, and several hemoglobin
components including HbA1c are separated by high performance liquid
chromatography. Each component is detected by a single wavelength
colorimeter, and the detected data is processed by a computer to
calculate the percentage content of HbA1c in the total
hemoglobin.
Specification
Model | LD-560 |
Method | HPLC |
Parameters | HbA1c, HbF, HbA1a, HbA1b, HbA0; HbA2 (optional) |
Test Speed | 90 seconds per test |
Reagents kit | Eluent A, Eluent B, Hemolysis |
Autoloader Capacity | 20 tubes |
Precision | CV<2% |
Dimension | 378(L) ×380(W)×510(H)mm |
Weight | 28kg |
Column Capaticy | 3000 Tests |
Report Range | 3%-18% |
Working Condition | Temperature: 15-30℃ Humidity:30%-85% |
Voltage | 100-240VA |
Frenquency | 50/60Hz |
Dimension | 610(L) ×435(W)×H425(H)mm |