Product Details
HbA1c Analyzer LD-100 Hemoglobin Testing Analyzer POCT One Step
Detection Fast Testing Analyzer
Intended Use
- It is intended for the determination of percent hemoglobin A1c (HbA1c) in human whole blood.
- It is a semi-automatic instrument used to analyze glycated
hemoglobin.
- It can be used to detect of HbA1c in medical units at all
levels.And it's very protable to carry, you can testing HbA1c
anytime.
- It has fully automatic system just need one step operation then the
test will be start.
Specifications
Model | LD-100 |
Sample Type | Whole blood |
Sample Volume | 4ul |
Test Time | Less than 3 minutes |
Data storage | 7000 |
Method | Fluorescent Immunoturbidimetry |
Operating Type | 1-Step Operation, Fully automatic system |
Measurement Range | HbA1c 4.0~15%(NGSP Unit) eAG 68-385mg/dl(mmol/l) |
Operating Condition | Temperature:15~35℃ Humidity:15~80%R.H |
Storage Condition | Analyzer:10-60℃; Test kit: 2~8℃ |
Voltage | 12VDC-3.33A |
Analytical wave-length | 510nm and 580nm. |
Certification | NGSP, IFCC |
PRINCIPLE
- Uses the (borate affinity) fluorescent immunoturbidimetry
- Push a mixture containing a fluorescent borate conjugate into an
integrated test cup and mix to dissolve sufficiently.
- The blood samples in the sample collector are then pushed into the
buffer and mixed to release the hemoglobin from the red blood
cells.
- The total concentration of hemoglobin depends on the initial
reduced fluorescence signal.
- Fluorescent borate conjugates bind to glycosylated hemoglobin, and
glycosylated hemoglobin is determined by detecting the decrease of
fluorescent active ingredients.
- The percentage of glycosylated hemoglobin in total hemoglobin can
be determined.
Feature
- Easy to operate, One Step Testing
- High Accuracy
- Cv<3%, Excellent Correlation With HPLC
- RFID For Batch Management
- Color touch screen
- External printer (Blue-tooth)
- USB Port, LIS Support
- One analyzer for testing without extra instrument
Reliable
IFCC calibrator | Normal | Abnormal | Normal | Abnormal |
Mean HbA1c% | 5.4 | 9.6 | 5.3 | 9.2 |
CV (%) | 2.5% | 1.9% | 2.4% | 1.8% |
Use Step
- Place cartridge in analyzer
- Push bead into reagents
- Insert sample into cartridge
Performance Parameters
- Accuracy: Is tested with the reference material as samples and the relative
deviation of the determination results of the analyzer should be
within the range of ±8%.
- Precision:Samples with a concentration of 4.0%-6.5% are detected, sample
coefficient of variation (CV) of the test results is not more than
3%.
- Linearity:In the range of 4%-16% of HbA1c, the linear correlation
coefficientrof the test results should be no less than 0.9900.
- Stability:Within 8 hours after the instrument is turned on and stable, the
relative deviation of the test results is not more than ±3%.
Certificated
Company Profile
Labnovation Technologies, Inc. is an original manufacturer
established in Shenzhen, China since 2001. We are dedicated in
developing and manufacturing IVD instruments and reagents including
HbA1c analyzer and reagents, rapid test reader & reagents,
hematology reagents, clinical chemistry reagents, electrolyte
reagents, wash solution and urine test reagents.
Labnovation products are developed in house, tested extensively in
clinics and passed strict quality controls. All of our products are
CE marked and approved by SFDA. In the past 15 years, our products
have been exported to 110 different countries and used in 8000
hospitals worldwide. We provide both ready-to-use final products
and bulk OEM reagents, including ready-to-use reagents,
concentrated, and powder.
With an investment of 10% of our annual sales into R&D, we
release new products every years. We win more and more domestic and international cooperation by
company reputation and products quality while we actively take part
in fairs like MEDICA,AACC, ARBA HEALTH and so on through which we
are well-known by the companies in this field.