Product Details
HbA1c Analyzer POCT One Step HbA1c Testing Analyzer LD-100
Hemoglobin Analyzer
Product application
This instrument is intended for the percent determination of
hemoglobin A1c (HbA1c) in human whole blood.
Specifications
Model | LD-100 |
Method | Boronate Affinity |
Operating Type | 1-Step, fully automatic |
Measurement Time | Less than 3 minutes |
Sample Type | Whole blood |
Measurement Range | HbA1c 4.0~15%(NGSP Unit) eAG 68-385mg/dl(mmol/l) |
Operating Condition | Temperature:15~35℃ Humidity:15~80%R.H |
Dimensions | 220(w)x150(L)x120(H)mm |
Shelf-Life of Test Kit | 9 months |
Weight | 800gs |
Storage Condition | Analyzer:10-60℃; Test kit: 2~8℃ |
Voltage | 12VDC-3.33A |
PRINCIPLE
HbA1c Analyzer uses the (borate affinity) fluorescent
immunoturbidimetry. Push a mixture containing a fluorescent borate
conjugate into an integrated test cup and mix to dissolve
sufficiently. The blood samples in the sample collector are then
pushed into the buffer and mixed to release the hemoglobin from the
red blood cells. The total concentration of hemoglobin depends on
the initial reduced fluorescence signal. Fluorescent borate
conjugates bind to glycosylated hemoglobin, and glycosylated
hemoglobin is determined by detecting the decrease of fluorescent
active ingredients. The percentage of glycosylated hemoglobin in
total hemoglobin can be determined
Performance Indicators
- Result reporting unit: Glycated hemoglobin (HbA1c) is reported by
default in IFCC units (mmol/mol) and NGSP units (%).
- Accuracy: The relative deviation of the determination results of
the analyzer should be within the range of ±8%.
- Precision: Samples with a concentration of 4.0%-6.5% are detected,
and the sample coefficient of variation (CV) of the test results is
not more than 3%.
- Linearity: In the range of 4%-16% of HbA1c, the linear correlation
coefficient r of the test results should be no less than 0.9900.
Feature
- One-step Operation
- Battery Integrated
- Portable, Compact
- RFID For Batch Management
- Color touch screen
- External printer (Blue-tooth)
- USB Port, LIS Support
Reliable
IFCC calibrator | Normal | Abnormal | Normal | Abnormal |
Mean HbA1c% | 5.4 | 9.6 | 5.3 | 9.2 |
CV (%) | 2.5% | 1.9% | 2.4% | 1.8% |
Use Step
- Place cartridge in analyzer
- Push bead into reagents
- Insert sample into cartridge
Certificated
Company Profile
Labnovation Technologies, Inc. is an original manufacturer
established in Shenzhen, China since 2001. We are dedicated in
developing and manufacturing IVD instruments and reagents including
HbA1c analyzer and reagents, rapid test reader & reagents,
hematology reagents, clinical chemistry reagents, electrolyte
reagents, wash solution and urine test reagents.
Labnovation products are developed in house, tested extensively in
clinics and passed strict quality controls. All of our products are
CE marked and approved by SFDA. In the past 15 years, our products
have been exported to 110 different countries and used in 8000
hospitals worldwide. We provide both ready-to-use final products
and bulk OEM reagents, including ready-to-use reagents,
concentrated, and powder.
With an investment of 10% of our annual sales into R&D, we
release new products every years. We win more and more domestic and international cooperation by
company reputation and products quality while we actively take part
in fairs like MEDICA,AACC, ARBA HEALTH and so on through which we
are well-known by the companies in this field.