LD-100 Hematology Detection Analyzer Medical One Step Hemoglobin A1c Analyzer

Brand Name:LABNOVATION
Certification:IFCC, NGSP, CE
Model Number:LD-100
Minimum Order Quantity:1
Delivery Time:10 work day
Payment Terms:D/A, D/P, T/T
Contact Now

Add to Cart

Active Member
Location: Shenzhen China
Address: 101 and 5th Floor, Building 1, No. 68,18th Road, Guangming Hi-Tech Park, Tangjia Community, Fenghuang Street, Guangming District, Shenzhen 518107 ,China
Supplier`s last login times: within 29 hours
Product Details Company Profile
Product Details

HbA1c Analyzer LD-100 Hematology Detection Analyzer Medical Use HbA1c Analyzer


Intended Use

This instrument is intended for the percent determination of hemoglobin A1c (HbA1c) in human whole blood.

Specifications

ModelLD-100
MethodBoronate Affinity
Operating Type1-Step, fully automatic
Measurement TimeLess than 3 minutes
Sample TypeWhole blood
Measurement RangeHbA1c 4.0~15%(NGSP Unit)
eAG 68-385mg/dl(mmol/l)
Operating ConditionTemperature:15~35℃
Humidity:15~80%R.H
Dimensions220(w)x150(L)x120(H)mm
Shelf-Life of Test Kit9 months
Weight800gs
Storage ConditionAnalyzer:10-60℃; Test kit: 2~8℃
Voltage12VDC-3.33A

PRINCIPLE
HbA1c Analyzer uses the (borate affinity) fluorescent immunoturbidimetry. Push a mixture containing a fluorescent borate conjugate into an integrated test cup and mix to dissolve sufficiently. The blood samples in the sample collector are then pushed into the buffer and mixed to release the hemoglobin from the red blood cells.


Performance Parameters

  • Accuracy:The relative deviation of the determination results of the analyzer should be within the range of ±8%.
  • Precision:Samples with a concentration of 4.0%-6.5% are detected, and the sample coefficient of variation (CV) of the test results is not more than 3%.
  • Linearity:In the range of 4%-16% of HbA1c, the linear correlation coefficient r of the test results should be no less than 0.9900.
  • Stability:Within 8 hours after the instrument is turned on and stable, the same normal sample is tested, and the relative deviation of the test results is not more than ±3%.

Feature

  • One-step Operation
  • Boronate Affinity
  • Battery Integrated
  • Portable, Compact
  • Cv<3%, Excellent Correlation With HPLC
  • RFID For Batch Management
  • Color touch screen
  • External printer (Blue-tooth)
  • USB Port, LIS Support

Reliable

IFCC calibratorNormalAbnormalNormalAbnormal
Mean HbA1c%5.49.65.39.2
CV (%)2.5%1.9%2.4%1.8%

Use Step

  • Place cartridge in analyzer
  • Push bead into reagents
  • Insert sample into cartridge


Certificated


China LD-100 Hematology Detection Analyzer Medical One Step Hemoglobin A1c Analyzer supplier