CE Fluorescence Immunoassay Test Triiodothyronine T4 Test Cassette

Brand Name:FIATEST
Certification:CE
Model Number:FI-T4-302
Minimum Order Quantity:N/A
Delivery Time:2-4 Weeks
Payment Terms:L/C, T/T
Contact Now

Add to Cart

Verified Supplier
Location: Hangzhou Zhejiang China
Address: #383, Qiaoxin Road, Xiasha Street, Qiantang Disctrict, Hangzhou, Zhejiang, P.R.China
Supplier`s last login times: within 48 hours
Product Details Company Profile
Product Details

Triiodothyronine T4 Test Cassette Hormone Marker CE, Determination of total Thyroxine (T4) in serum or plasma


T4 Test Cassette is intended for in vitro quantitative determination of total Thyroxine (T4) in serum or plasma. Measurement of T4 is used as an aid to assessment of thyroid function.The test result is calculated by Fluorescence Immunoassay Analyzer.


Product Name:Triiodothyronine T4 Test Cassette Hormone Marker CE, Determination Of Total Thyroxine (T4) In Serum Or Plasma
Format:Cassette
Certificate:CE
Pack:10T/25T
Shelf Life:2 Years
Precision:C.V. ≤ 15%.
Principle:Fluorescence Immunoassay
Specimen:S/P
Reading Time:15 Minutes
Storage Temperature:4-30℃
Cut-Off:65-155nmol/L

PRODUCT FEATURES


【PERFORMANCE CHARACTERISTICS】

1. Accuracy

The test deviation is ≤ ±15%.

2. Assay Range and Detection Limit

Assay Range: 12.87~300nmol/L.

Detection Limit (Analytical Sensitivity): 12.87nmol/L.

3. Precision

C.V. ≤ 15%.


Fast Result (10 minutes)

Simple Operation (Less Training Required)

Objective (Results Read by Analyzer)

Stringent Quality Control Insure High Accuracy

User-friendly (Simple Plug & Play Operation)

High Efficiency (Both STAT Testing and Batch Testing)



【DIRECTIONS FOR USE】

Refer to FIATESTTM Fluorescence Immunoassay Analyzer Operation Manual for the complete instructions on use of the Test. The test should be in room temperature.


Allow the test cassette, specimen, buffer and/or controls to reach room temperature (15-30°C) prior to testing.

  1. Turn on the Analyzer power. Then according to the need, select “standard test” or “Quick test” mode.
  2. Take out the ID card and insert it into the Analyzer ID Card Slot.
  3. Pipette 900 μl of reagent 1 dilution buffer into the reagent 2 tube and dissolve the lyophilized powder thoroughly.

Note:After dissolution,reagent 2 could be stored at 2-8℃ for 30 days or stored at -20℃ for long period.

  1. Draw 75ul of the dissolved reagent 2(step 3) into 0.6ml bufffer tube.
  2. Pipette 75 μl of serum or plasma into the buffer tube, mix the specimen and the dissolved reagent 2 well.
  3. Leave the mixture for reaction of 5 minutes, then pipette 75 μl diluted sample into the sample well of the cassette. Start the timer at the same time.
  4. There are two test modes for FIATESTTM Fluorescence Immunoassay Analyzer, Standard Test mode and Quick Test mode. Please refer to the user manual of FIATESTTM Fluorescence Immunoassay Analyzer for details.

“Quick test” mode: After 15 minutes of adding sample,Insert the test cassette into the Analyzer, click “QUICK TEST”, fill the test information and click "NEW TEST" immediately. The Analyzer will automatically give the test result after a few seconds.

“Standard test” mode: Insert the test cassette into the Analyzer immediately after adding specimen, click “STANDARD TEST”, fill the test information and click "NEW TEST" at the same time, The Analyzer will automatically countdown 15 minutes. After the countdown, the Analyzer will give the result at once.


【INTERPRETATION OF RESULTS】

Results read by the Fluorescence Immunoassay Analyzer and display the result on the screen.

Cat.No.ProductSpecimenFormatKit SizeCE Status
FI-T4-302T4 Test CassetteSerum/PlasmaCassette10T/25 TCE
China CE Fluorescence Immunoassay Test Triiodothyronine T4 Test Cassette supplier

CE Fluorescence Immunoassay Test Triiodothyronine T4 Test Cassette

Inquiry Cart 0