Add to Cart
Eight-Channel Optical Power Meter With Display Optical Signal Monitoring Device
An eight-channel optical power meter with display is an instrument specifically designed to measure optical power. It has multiple channels, can measure multiple optical signals simultaneously, and is equipped with a display screen to visually display the measurement results.
Product characteristics
Multi-channel design: The eight-channel design allows for
simultaneous monitoring of multiple optical signals, increasing
test efficiency.
High-precision measurement: Advanced detection circuits and
technologies are used to ensure the accuracy and reliability of
measurement results.
Color LCD display: Equipped with a 3.5 inch or similar size color
LCD screen, the display interface is clear and easy to operate.
Programmable control: Support USB and RS232 and other communication
interfaces, can be controlled by the program to complete all the
test process, easy to build automatic measurement and control
system.
Independent working wavelength setting: Each channel can be
independently set working wavelength to meet different test
requirements.
Application field
Eight-channel optical power meter with display screen is widely
used in communication, optoelectronics, scientific research and
other fields, especially in the test and maintenance of optical
fiber communication system plays an important role. It can be used
to measure parameters such as optical fiber attenuation, optical
power distribution, and optical device performance, helping
engineers and technicians quickly and accurately locate problems
and take appropriate measures.
Technical specification
Model # | U8201 | U8202 | U8204 | U820XH |
Number of Channels | 1 | 2 | 4 | X |
Power Range | +10 ~ -65dBm | +25 ~ -50dBm | ||
Wavelength Range | 850 ~ 1700 nm | |||
Calibration Wavelength | 850/980/1310/1490/1550/1610 nm | |||
Application Fiber Type | Standard SM and MM up to 62.5 um core size | |||
Absolute Uncertainty (accuracy) at reference condition | ± 4% (1200 nm ~ 1610 nm) | |||
Relative Uncertainty (accuracy) at reference condition | < 0.02 dB Typical | |||
Linearity (power) | ≤ ± 0.06 dB (1200 nm ~ 1610 nm, 0~ -50dBm) | |||
Return Loss | > 40 dB | |||
Operation Temperature | 0 ~ +40℃ | |||
Storage Temperature | -30 ~ +80℃ | |||
Recalibration Period | 2 years | |||
Dimensions | 245 mm W, 105 mm H, 320 mm D | |||
Weight | 2.0 kg |
Precautions for use
Correct connection: Before using the optical power meter, ensure
that the optical power meter is properly connected to the optical
fiber or optical device to be measured to avoid measurement errors
caused by loose interfaces or improper connection.
Calibration: Calibrate the optical power meter periodically to
ensure the accuracy of the measurement results. The specific
calibration method can refer to the instrument manual or consult
the manufacturer's technical personnel.
Dust and moisture: Keep the working environment of the optical
power meter clean and dry to avoid dust and moisture damage to the
instrument.
Battery power: Check the battery power to ensure that the battery
power is sufficient during the test to avoid measurement
interruption or data loss due to insufficient power.