- The 2019-nCoV IgG/IgM Rapid Test Cassette (Whole
Blood/Serum/Plasma) can be performed using whole blood (from
venipuncture or fingerstick), serum or plasma.
- To collect Fingerstick Whole Blood Specimens:
- Wash the patient’s hand with soap and warm water or clean with an
alcohol pad. Allow to dry.
- Massage the hand without touching the puncture site by rubbing down
the hand towards the fingertip of the middle or ring finger.
- Puncture the skin with a lancet. Wipe away the first sign of blood.
- Gently rub the hand from wrist to palm to finger to form a rounded
drop of blood
DIRECTIONS FOR USE
Allow the test, specimen, buffer and/or controls to reach room
temperature (15- 30°C) prior to testing.
1. Remove the test cassette from the foil pouch and use it within
one hour. Best results will be obtained if the test is performed
immediately after opening the foil pouch.
2. Place the cassette on a clean and level surface. For Serum or
Plasma specimen:
To use a dropper: Hold the dropper vertically, draw the specimen to
the fill line (approximately 10uL), and transfer the specimen to
the specimen well (S), then add 2 drops of buffer (approximately
80uL), and start the timer.
To use a pipette: To transfer 10 uL of specimen to the specimen
well(S), then add 2 drops of buffer (approximately 80 uL), and
start the timer
For Venipuncture Whole Blood specimen:
To use a dropper: Hold the dropper vertically, draw the specimen
about 1 cm above the fill line and transfer 1 full drop (approx.
20μL) of specimen to the sample well(S). Then add 2 drops of buffer
(approximately 80 uL) and start the timer.
To use a pipette: To transfer 20 uL of whole blood to the specimen
well(S), then add 2 drops of buffer (approximately 80 uL), and
start the timer
For Fingerstick Whole Blood specimen:
To use a dropper: Hold the dropper vertically, draw the specimen
about 1 cm above the fill line and transfer 1 full drop (approx.
20μL) of specimen to the sample well(S). Then add 2 drops of buffer
(approximately 80uL) and start the timer.
To use a capillary tube: Fill the capillary tube and transfer
approximately 20uL of fingerstick whole blood specimen to the
specimen well (S) of test cassette, then add 2 drops of buffer
(approximately 80 L) and start the timer. See illustration below.
3. Wait for the colored line(s) to appear. Read results at 10
minutes. Do not interpret the result after 20 minutes.
INTERPRETATION OF RESULTS
IgG POSITIVE:* Two colored lines appear. One colored line should
always appear in the control line region (C) and another line
should be in the IgG line region.
IgM POSITIVE:* Two colored lines appear. One colored line should
always appear in the control line region (C) and another line
should be in the IgM line region.
IgG and IgM POSITIVE:* Three colored lines appear. One colored line
should always appear in the control line region (C) and two test
lines should be in the IgG line region and IgM line region.
*NOTE: The intensity of the color in the test line regions may vary
depending on the concentration of 2019-nCoV antibodies present in
the specimen. Therefore, any shade of color in the test line region
should be considered positive.
NEGATIVE: One colored line appears in the control line region (C).
No line appears in the IgG region and IgM region.
INVALID: Control line fails to appear. Insufficient specimen volume
or incorrect procedural techniques are the most likely reasons for
control line failure. Review the procedure and repeat the test with
a new test. If the problem persists, discontinue using the test kit
immediately and contact your local distributor.
QUALITY CONTROL
Internal procedural controls are included in the test. A colored
line appearing in the control region (C) is an internal procedural
control. It confirms sufficient specimen volume and correct
procedural technique. Control standards are not supplied with this
kit; however, it is recommended that positive and negative controls
be tested as a good laboratory practice to confirm the test
procedure and to verify proper test performance