In Line Eluent Generation Ion Chromatograph Instruments With Conductivity Detector

Brand Name:Wayeal
Certification:ISO, CE
Model Number:IC6200
Minimum Order Quantity:1
Delivery Time:30 days
Payment Terms:L/C, T/T
Contact Now

Add to Cart

Verified Supplier
Location: Hefei Anhui China
Address: No. 8, Wenqu Road, High-tech Zone, Hefei, Anhui
Supplier`s last login times: within 48 hours
Product Details Company Profile
Product Details

In-line Eluent Generation Ion Chromatograph Instruments with Conductivity Detector


Introduction of IC6200 Ion Chromatography Instrument

IC analysis uses a High Performance Liquid Chromatography (HPLC) system equipped with a conductivity detector to separate anions or cations. The ions in a sample are moved through the system in solution and then the various ions are separated in a column with an ion-exchange resin. The resin’s surface layer has limited ion-exchange sites that attract either anions or cations. A conductivity detector at the end of the column measures the quantity of exiting ions, and from this data a chromatogram is produced.


Features of Ion Chromatography Instrument

1. Intergrated apperance design, high brightness LCD full touch GUI controller
2. High degree of automation "One Key Rinse" and "One Key Maintenance" automatic maintenance function.
3. Higher intergration: Support optional parts upgrade to 2D system and multi-detector combination.
4. Full PEEK pump and electric injection valve.

5. Powerful and convenient chromatpgraphy workstation

6. High precision temperatur-controlled conductivity detector.


Technical Data of Ion Chromatography Instrument

In Line Eluent Generation Ion Chromatograph Instruments With Conductivity Detector
System Parameter
Qualitative Repeatability≤0.5%
Quantitative Repeatability≤1%
Pump System
Flow Rate(0.001~10.000)mL/min
Pump Head Pressure Resistance(0 ~ 35) MPa
Flow Rate Setting Inaccuracy<0.1%
Flow Stability<0.1%
Pressure Display Accuracy0.01Mpa
Pressure Pulsation<0.5%
Conductivity Detector
Baseline Noise≤0.1%FS
Baseline Drift≤0.5%FS
Minimum Detectable Concentration (Cl)≤0.0002μg/mL
Minimum Detection Concentration (Li)≤0.0001μg/mL
Conductivity Cell Volume≤ 0.6μL
Conductivity Detection Range0-15000μS/cm
Detector Resolution0.00238nS/cm
Conductivity Cell Temperature Setting Error±0.01°C
Resolution of Conductivity Cell Temperature Display0.001°C
Suppressor
Constant Current Source Range0-500mA in 0.1mA increments
Column Oven
Temperature RangeRoom temperature +5°C~85°C
Control Temperature Stability≤ 0.1°C
Autosampler
Number of Sample Vials≥ 120 (1.5mL standard sample bottle)
Repeatability

Full loop injection < 0.3% RSD,

Partial injection < 0.5% RSD,

Non-destructive injection < 1 % RSD (injection volume > 5uL)

Syringe SizeStandard: 500μL; optional; 250μL, 1000μL, 2500μL
Loop SizeStandard: 100μL; optional: 20μL, 50μL, 200μL
Linearity>0.999
Cross-contamination<0.01%
Eluent Generator
Flow Rate Range0.001 to 3.000mL/min
Eluent Concentration RangeKOH, MSA-0.1 to 100mM
Concentration Accuracy0.1mM
Original Concentration of Electrolyte25% KOH/MSA
Volume of Electrolyte1000mL
Maximum Working Pressure21MPa (3000psi)
Gradient Accuracy1.5%

Industry Application

Detection of inorganic anions, iodides, disinfection by-products, adsorbable organohalogens, glyphosate, soluble cations and hexavalent chromium in environmental water quality.