Add to Cart
450 ℃Maximum operating temperatureLiquid Injection Laboratory Analysis FID ECD Gas Chromatography Analyzer
Product Introduction
The GC6000 series gas chromatographs achieve high automation, high efficiency and high precision. A wide range of detectors (FID, FPD, ECD, TCD, RID) are available as options. Three independent circuits, a gas system and three signal outputs are available simultaneously.
Product Parameter
Liquid Injection Laboratory Analysis FID ECD Gas Chromatography Analyzer | |
Laboratory Environment | |
Supply Voltage | 220V±10% |
Temperature | 18-28℃ |
Humidity | 40-70% |
Whole Machine Parameters | |
Qualitative Repeatability | ≤0.1% |
Quantitative Repeatability | ≤1.5% |
Baseline Noise | 0.01pA |
Baseline Drift | 0.1pA/30min |
Inlet Port | |
Maximum Operating Temperature | 400℃ |
Flow Control Mode | EFC |
Control Accuracy | 0.001psi |
Maximum Shunt Ratio | 999:1 |
Optional Carrier Gas Types | N2, H2, He, Ar |
Sampler | |
Linearity | 0.999 |
Cross-contamination | 0.01% |
Number of Injection Bits | 16 bits |
Number of Reagent Bits | 3 bits |
Column Oven | |
Above Room Temperature | +4℃- 450℃ |
Programmed Temperature Rise Rate | 40℃/min |
Program temperature rise repeatability | <0.5% |
Minimum Adjustment Amount | ≦0.01℃ |
Temperature Control Accuracy | ±0.05℃ |
Maximum run time | 999.9min |
Detector (hydrogen flame (FID) detector) | |
Maximum operating temperature | 450 ℃ |
Minimum detection limit | 2.5pg / s |
Linear range | ≥ 106 |
Software
1. Self-developed, with software copyright
2. Workstation interface is simple, intuitive, convenient operation
process, with a comprehensive audit trail and rights management
functions
3. Graphical full counter-control interface, providing suitable
real-time analysis of the parameters and analysis results, online
modification and acquisition of the working parameters of the
components, automatic and rapid data acquisition and
post-processing, with instrument-related data and operating
conditions traceability function, convenient troubleshooting.
4. Realize the effective integration and control of system
components, provide chromatographic detectors, autosampler and
other upgrade components for the company's products can be
seamlessly added
Product Application
Application of gas chromatography in health inspection
Air, polycyclic aromatic hydrocarbons, benzene, toluene,
benzo(a)ratio, etc.
Residual organochlorine and organophosphorus pesticides in crops,
etc.
Food additives such as benzoic acid.
Analysis of biological materials such as body fluids and tissues
e.g. amino acids, fatty acids, vitamins, etc.