Shrimp Tissues Vpa Vibrio Parahaemolyticus Real Time PCR detection Kit

Brand Name:Biokey
Certification:ISO 13485
Model Number:BIK-Q-W007
Minimum Order Quantity:We can produce liquid and lyophilized kits
Delivery Time:Depending on order quantity
Payment Terms:L/C, T/T, Western Union
Contact Now

Add to Cart

Verified Supplier
Location: Guangzhou
Address: Add:No.2, Ruitai Rd, Guangzhou Hi-Tech Industrial Development Zone, 510670, CHINA
Supplier`s last login times: within 29 hours
Product Details Company Profile
Product Details

Vibrio parahaemolyticus kit Nucleic Acid PCR Detection Kit

(PCR-Fluorescent Probe Method)

Intended use:

This kit is suitable for the specific detection of pathogenic vibrio parahaemolyticus (Vpa) in bait, water samples, shrimp seedlings, shrimp tissues and processed products.


Principle:


This kit uses a specific fragment of Vibrio parahaemolyticus (Vpa) to design a primer probe probe, which can specifically bind to a SECTION of DNA template in the middle of the primer amplification region. During the PCR extension process, the exonuclide activity of Taq enzyme cuts the 5 '-terminal fluorophore from the probe and makes it free in the reaction system. Thus, the 3 '-end quenched group was separated, and the fluorescence was emitted, which was detected by fluorescence PCR instrument to achieve the detection of Vibrio parahaemolyticus nucleic acid (Vpa) in the completely closed reaction system.

Main components:

ComponentSpecification
PCR reaction liquid24 tubes
Positive control1 tube
Negative control1 tube
Note: components in different batches of kits should not be interchangeable.


Sample requirements:

1. Sample type

Shrimp seedling, shrimp tissue, water body, bait, processed products, etc


2. Sample collection and transportation

Sample collection and transportation shall be carried out in accordance with NY/T 541. Samples shall be transported to the laboratory as soon as possible under refrigerated conditions to avoid repeated freeze-thaw.


3. Sample preservation

It can be stored for 24 hours at 2℃~8℃, and can be stored for a long time at -20℃.



Using Micgene 242/244/244IVD and ABI QuantStudio5:

procedures are set as below:

Steps

Cycle number

TemperatureTime
1195℃3min
24595℃10s
60℃

30sCollect the fluorescent

FAM was selected for detection channel.

The results of interpretation

ResultsCt Value
PositiveCt≤40
NegativeNo Ct value or Ct value≥45
Gray area40<Ct<45

Note: The result is judged to be grey area, which needs to be rechecked. If the recheck result is Ct value < 45 is considered as positive, and no Ct value or Ct value ≥45 is considered as negative.

Result analysis:
Automatic analysis using instrumentation

Quality control:
Negative control: No obvious amplification curve of FAM detection channel;
Positive control: FAM channel showed obvious amplification curve, Ct value ≤30;

The above conditions are met simultaneously in the same experiment; otherwise, the experiment is invalid and re-testing is required.