Add to Cart
Altitude & Pressure Test Chamber for Battery Testing IEC 61960 150L
Product information:
The altitude and pressure test chamber is widely utilized in various research and application fields, including biochemistry, chemical and pharmaceutical industries, healthcare, agricultural research, new energy, and electronics. It is designed for rapid and efficient drying of powders, baking, and drying heat-sensitive, easily decomposable, or oxidizable substances, as well as materials with complex compositions. The chamber complies with industry standards such as UN 38.3, UL 1642, and IEC 61960.
Structure:
1. The chamber is made of high-quality cold-rolled steel plate with an electrostatic powder coating on the surface, providing a hard and durable coating with excellent rust resistance.
2. The working chamber is made of high-quality brushed stainless steel plate with a thickness of 5mm.
3. Reinforcement treatment is applied between the chamber and the working chamber.
4. The chamber door is constructed with a double-layer tempered glass structure, allowing a clear view of the items inside.
A rubber sealing ring is installed between the working chamber and the glass door to ensure a high level of vacuum inside the chamber.
Technical parameter:
Internal Dimensions | 500mm (W) × 500mm (D) × 600mm (H) |
Approximate External Dimensions | 700mm (W) × 700mm (D) × 1650mm (H) (subject to actual product) |
Control Panel | Located on the machine door |
Door Opening Method | Single door, opens from right to left |
Sealing Material | Silicone strip |
Chamber Structure | High-pressure product laboratory with a pressure display panel |
Pressurization Pump Installation | Installed at the lower part of the machine |
Internal Chamber Material | SUS304# stainless steel industrial plate, 5.0mm thick, reinforced to prevent deformation under pressure |
External Chamber Material | 1.2mm thick cold-rolled steel plate |
Operating Environment | temperature: 5 to 25℃ humidity: 35% to 75% |
Vacuum Range | Atmospheric pressure to an altitude of 8000m (vacuum level can be set as needed) |
Vacuum Leakage Rate | 0.5 KPA/hour (automatic compensation) |
Pressure Fluctuation | Less than 3% |
Evacuation Time | Approximately 20 minutes from atmospheric pressure to an altitude of 8000m |
Pressure Resolution | 0.01 KPA (altitude display accurate to meters) |
Power Supply | AC 220V 50Hz 2KW |
Internal Power Input | Equipped with an aviation plug and two/three-pin sockets, allowing for powered testing inside the chamber |
Sealing and vacuum display | 1.This machine uses a high-density silicone sealing strip at the door for excellent overall sealing performance. 2.A 7-inch custom touch-screen digital vacuum gauge is used to display the vacuum level inside the chamber and the negative pressure gauge. 3. Vacuum Sensor: Electronic pressure transmitter. 4. Vacuum Gauge Display Range: 0 KPa to 120 KPa. 5. Pressurization Control Method: Automatically controlled—stops when reaching the upper limit, starts when below the lower limit, cycling to maintain optimal high-pressure conditions. 6. Pressurization Pump Configuration: Equipped with one pressurization pump. 7. Pressure Relief & Exhaust System: Requires venting to the outside (automatic or manual). |