Endodontic wireless endo motor
General Information
1. ZR AUTO MOTOR is a compact device combining endodontic motor
with micorprocessor control for working with rotating Ni-Ti
Instruments.
2. The device is designed for us in dentistry(endodontics) and can
be used only in hospitals by medical specialists licensed to
practice dentisty. The manufacturer is not responsible for the
device misuse.
3. The list of programs endo motor
Programs | Mode of rotation | Name of the file system, under which adapted the program |
1,2,3,4 | programmable | |
5 | ||
6 | Forward 170° and reverse 50° | |
7 | Reverse 150° and forward 30° | |
8 | Reverse 170 °and forward 50° | |
9 | Continuous forward rotation normally, but when the load on the file exceeds the set limit, the file automatically starts alternately between forward 180° and 90° reverse rotation |
Instrument constantly rotates clockwise or counterclockwise,
respectively
Instruments alternately rotates clockwise and counterclockwise
with clearly fixed speed and rotation angle in both direction: with
a big rotation angle in both directions:with a big rotation angle
when moving forward(clockwise or counterclockwise, respectively)
and with a small when moving back
4. Safety measures and warnings
4.1 Use the device only with ZR AUTO MOTOR( Endodontic wireless
endo motor) authentic accessories
4.2 Do not dismount and do not change the device contstruction
anyhow. In the case of the device integrity damage warranty is
considered to be invalid. The device power source should be changed
by authorized service departments only.
4.3 Avoid any liquid ingress into the device case.
4.4 Do not use the device close to inflatmmables. The device is not
operational in the presence of inflammable anesthetic mixtures with
air, oxygen or nitorgen oxides.
4.5 Use only sterile and disinfected device components. The device
should be sterilized
and disinfected directly before the rst device use and also after
each patient in order
to avoid cross infection (for more details see part7 “Cleaning,
Disinfection and Sterilization”).
4.6 Be sure to lubricate and sterilize the micromotor head before
its use after previous patient.
Avoid lubricant ingress into the micromotor expanded parts.
4.7 Use cofferdam when working in patient’s oral cavity.
4.8 In case of any abnormality in the endomotor functioning (noise,
shakiness, vibrations),
stop using the device immediately and try to find the cause of
malfunction, basing on the
information represented in the part 8.Carefully follow
manufacturer’s instructions con-
cerning use of the endodontic instruments and avoid use of worn-out
and damaged files
4.9 Do not use instruments of “continuous” rotation in “reciprocal”
movement and vice versa.
4.10 Do not use curved, deformed or not corresponding to the ISO
file. Use of such a file may
result in injury.
4.11 Do not start the micromotor until the file is securely fixed
in the clutch head.
4.12 Avoid inadvertent contact with rotating micromotor parts
and/or rotating file
4.13 The device(Endodontic wireless endo motor) normally operates
at temperatures 10-35°С, relative air humidity not more than
80%, atmosphere pressure (101±3) kPa. Any violation of the pointed
restrictions may lead
to the device misoperation.