GENERAL DESCRIPTION
TMTECK Centrifuge Tubes are used to monitor the concentration of
magnetic particles and the level of contamination in fluorescent
and visible baths.
DAILY INSTRUCTIONS (INCLUDING NEW BATH)
1. Let pump motor run for several minutes to agitate the suspension
2. Flow the bath mixture through hose and nozzle for a few moments
to clear hose.
3. Fill the centrifuge tube to the 100 ml line.
4. Place the tube in the stand at a location free from vibration
and let stand for 30 minutes for a water bath and 60 minutes for an
oil bath to allow particles to settle out.
The gravity settling method applies to either oil or water
suspension. In hot weather the water bath should be checked more
often as is it more volatile than oil. Therefore, as water is lost
by evaporation, it must be replaced.
The settled particles (measured in ml) in the bottom of the tube
indicate the amount of magnetic particles in suspension. A UV
light, such as the MPXL Portable Black Light, must be used for
fluorescent particles.
Do not include dirt particles in your centrifuge tube readings.
They usually settle on the top of the magnetic particles.
Dirt will not fluoresce under a black light. In visible particles,
the appearance of dirt is very different than that of the
particles. Dirt will be coarser and irregular in shape. See
illustrations on page 3 for recommended settling volume.
BATH MAINTENANCE TIPS
To maintain proper bath suspension during inspection requires that
it be agitated prior to and while the bath is in use. The agitator
pipe should be removed and thoroughly cleaned monthly or more
often, if needed. Also, check area where sump screen connects to
tank, clean and remove any foreign material that may restrict flow.
Constant use of the bath requires a daily check for evaporation of
oil or water, loss of particles due to carry off and contamination.
Eventually the bath will become so contaminated by dirt, lint, oil
or other foreign material that efficient formation of indications
will become impossible. Contamination can be checked by noting the
amount of foreign material that settles out with the particles in
the centrifuge tube. Covering equipment, when not in use, will
reduce contamination and evaporation.
SPECIFICATIONS COMPLIANCE
- ASTM E709-08 (Sections 20.6.1 & X5)
- ASTM E1444/E1444M-12 (Section 7.2.1)
- BPVC (Section V, Article 7: T-765)