Add to Cart
>
Power supply | AC(220±10%)V,50Hz |
Heating device | (1) The furnace is made of SiC materials; its power is 600 W (2) Heating power is adjustable in the range of 0 W~600W |
Sample temperature rising rate | 1℃~12℃ per minute, manually controlled |
Electric stirring device | (1) Stirring motor: 45TYZ Motor (2) Driving device: by a soft shaft (3) Stirring blade specification: 8 mm×40 mm |
Stirring speed | As per the requirement of GB/T261-2008 (1) 90 RPM~120 RPM for procedure A (2) 250±10 RPM for procedure B |
Standard oil cup | As per the requirement of GB/T261-2008; (1) Inner diameter: 50.7 mm~50.8 mm (2) Depth: 55.7 mm~56.0 mm (3) Depth of marked line in the oil cup: 33.9 mm~34.3 mm (4) Oil sample cubage: about 70 ml |
Flame application device | (1) Gas source: Coal gas (or other civilian combustible gas) (2) Aperture of flame application device: 0.8 mm |
Thermometer | Rod type Mercury Thermometer, as per GB/T261-2008; (1) Temperature range: -5℃~110℃, Each scale division is 0.5 ℃ (2) Temperature range: 20℃~150℃, Each scale division is 1.0 ℃ (3) Temperature range: 90℃~370℃, Each scale division is 2 ℃ |
Ambient temperature | ≤35℃ |
Relative humidity | ≤85% |
Total power consumption | not more than 650 W |
No. | Name | Unit | Quantity | Remark |
1 | Main unit | Set | 1 | |
2 | Thermometer -5℃~110℃, each scale division is 0.5℃; | Piece | 1 | As per GB/T 261-2008 |
3 | Thermometer 20℃~150℃, each scale division is 1.0℃; | Piece | 1 | |
4 | Thermometer 90℃~370℃, each scale division is 2℃; | Piece | 1 | |
5 | Soft shaft | Piece | 1 | |
6 | Silica gel tube φ3 0.5 m | Piece | 3 | 1 is spare |
7 | Thermometer adapter | Set | 1 | |
8 | Thermometer liner band (cut suitable length of liner band to wrap
the lower end of the thermometer, so that the thermometer is just
inserted into the thermometer adaptor) | Piece | 5 |
Package Photos