Add to Cart
The TFOUT Tester adopts a full-automatic control system. The
operator can automatically test and print the results as long as
the sample is put into the instrument. The whole process does not
require manual conduct.
The principle is to put the sample, distilled water and copper
catalyst coil into the glass tube together, and then put it into
the oxygen bomb. The oxygen bomb is filled with oxygen at room
temperature and put into the oil bath at the specified temperature.
The oxygen bomb needs to form an angle of 30 degrees with the
horizontal plane, and the rotation speed of the oxygen bomb needs
to reach 100 R / min. When the specified pressure drop is reached,
record the time and stop the test as the oxidation stability of the
sample.
Feature
Applicable standards | ASTM D4742 and SH/T 0074 |
Instrument structure | all stainless steel structure of floor type oxygen bomb (except angle iron frame). |
Temperature control temperature | 0 ~ 160 ºC adjustable, temperature control accuracy 0.1 ºC. |
Display | large screen LCD |
Display content | Windows interface: 1.Control temperature and set temperature; 2.Time / pressure log curve (four channels are displayed simultaneously); 3.Test results and average values; 4. Fault prompt, etc. |
Auxiliary functions | a. Automatic oxygen charging and leakage detection; b. Overtemperature alarm function; c. Liquid level alarm function. |
Pressure detection accuracy | ± 0.1% |
Test data storage | 100 groups |
Power consumption | 2.6KW |