Add to Cart
Respiratory Anesthesia Test Connection System
Product introduction:
The system consists of two-way airflow generator, moisture
generator, air conveying system, weighing system and flow meter
1. The bidirectional air flow generator is a mechanically driven
piston to produce a sinusoidal flow.
2. The moisture generator (Hg) consists of three parts: a heating
water bath (through which air is emitted in both directions), a
rigid cylindrical gas storage container (with a maximum volume of
7L and a diameter of about 150mm, containing a 2L air storage bag),
and an insulation box (including water bath, gas storage container
and heating source).
3. The air delivery system consists of a T-piece with an inner
diameter greater than 15 mm and an exhaust pipe at least 200 mm
long.
4. The accuracy of the weighing instrument is higher than ± 0.1mg
in the mass range to be measured.
5. Flow meter with an accuracy of at least 5% of the reading.