Chemiluminescence Nitrogen Analyzer for Solid Liquid Gas Samples 0.1-10000mg/L

Brand Name:Shandong Shengtai Instrument Co., Ltd.
Certification:SH/T 0657 ASTM D5762-2017 SH/T0704 ASTM D4629 GB/T 17674
Model Number:SH708
Minimum Order Quantity:1
Delivery Time:7-15 work days
Payment Terms:T/T
Contact Now

Add to Cart

Verified Supplier
Location: Jinan Shandong China
Address: Times Headquarters Base, Tianqiao District, Jinan City
Supplier`s last login times: within 13 hours
Product Details Company Profile
Product Details

SH708 Chemiluminescence nitrogen analyzer

The SH708 chemiluminescence nitrogen analyzer uses chemiluminescence method to determine the total nitrogen content, which improves the ability to resist impurity interference, avoids the complicated operation of the titration cell caused by the electric method, and greatly improves the sensitivity of the instrument. The key components of the system are imported, ensuring reliable performance of the entire machine. Suitable for determining the total nitrogen content of paraffin oil, diesel, gasoline, lubricating oil, fuel oil, liquefied gas, and natural gas, as well as other oil products, chemical raw materials, and finished products. It follows the standard SH/T 0657 for the determination of trace nitrogen in liquid petroleum hydrocarbons (oxidation combustion and chemiluminescence method, ASTM D5762-2017, SH/T0704, ASTM D4629, GB/T 17674 for the determination of nitrogen content in crude oil by boat injection chemiluminescence method).

Performance characteristics

Using pneumatic support to support the instrument cover, the internal structure of the instrument cover can be easily seen when opened.

Low and low content peaks have better shape and repeatability.

The total nitrogen content in the sample is rapidly measured by chemiluminescence method.

The application of high-precision data acquisition and computer technology provides reliable guarantees for data collection, control, and processing.

Technical parameters:

● Basic parameters: Sample types: liquid, solid, and gas

● Measurement method: chemiluminescence method

● Sample injection volume:

Solid sample: 1-20mg

Liquid sample: 5-20 μ L

Gas sample: 1-5mL

● Measurement range: 0.1-10000mg/L

● Measurement accuracy:

Concentration value (ppm)Injection volume (μ L)RSD(%)
0.12025
51010
50105
100103
5000103

● Temperature control range: Room temperature ~ 1050 ℃

● Temperature control accuracy: ± 3 ℃

● Gas source requirements:

High purity argon gas: purity above 99.995%

High purity oxygen: purity above 99.995%

● Power supply: AC220V ± 22V 50Hz ± 0.5Hz 1500 W

● Dimensions: Host: 305 (W) × 460 (D) × 440 (H) mm

● Exterior weight: 20kg

● Temperature control size: 550 (W) × 460 (D) × 440 (H) mm

● Temperature controlled weight: 40kg

Basic parametersSample types: liquid, solid, and gas
Measurement methodchemiluminescence method
Sample injection volume

Solid sample: 1-20mg

Liquid sample: 5-20 μ L

Gas sample: 1-5mL

Measurement range0.1-10000mg/L
Temp control rangeRoom temp ~ 1050 ℃
Temp control accuracy± 3 ℃
Gas source requirements

High purity argon gas: purity above 99.995%

High purity oxygen: purity above 99.995%

Power supplyAC220V ± 22V 50Hz ± 0.5Hz 1500 W
Host size305 (W) × 460 (D) × 440 (H) mm
Host weight20kg
Temperature control size550 (W) × 460 (D) × 440 (H) mm
Temperature controlled weight40kg

Note: The appearance pictures, dimensions, volume, and weight of the instrument are for reference only. If there are any updates to the instrument, there may be slight changes without prior notice.

Requirements for instrument completeness: host (data acquisition, control system), cracking furnace, automatic sampler (gas, liquid), computer operation software, etc.

Other optional items: computer, solid sampler, gas sampler (optional).


Shandong Shengtai Instrument Co., Ltd. provides the following quality assurance for the instruments sold to you:

----The provided instrument materials are brand new, meet national quality standards, and have a manufacturer's certificate of conformity;

----The materials and main components provided meet the technical requirements specified in the technical documentation;

----The overall quality assurance period of the equipment is one year (excluding normal wear and tear of vulnerable parts).

----We are responsible for free maintenance of any instrument quality issues that occur during the warranty period. Due to equipment malfunction caused by the user's responsibility, we are responsible for repairing and charging reasonable fees.

----Lifetime discounted supply of parts for equipment, and lifetime maintenance and repair of the entire machine.

----After the expiration of the warranty period, if the user needs maintenance and technical services, we will only charge a cost fee.

NumberNameQuantityUnitSpecsNote
1Host1Unit
2Injector1Unit
3Quartz tube1PieceInstall and carry
4Standard sample4Box
5Quartz elbow1Piece
6Membrane dryer1PieceImported from Japan
7Communication cable1Root
8Power line2Root
9Cracking furnace power supply1Root
10PTFE8MetreΦ3
11Silicone hose20cmΦ3
12Silicone hose20cmΦ5
13Large clamp1Piece
14Warranty Card Certificate of Conformity1Piece
15Instructions1Copy
16Half Moon Ceramic Cup4Piece
17Syringe2Piece10ul
18Fuse2Piece2A
19Pneumatic pressure cap4Piece
20Thermocouple1PieceS-type platinum rhodium platinum
21Sample injection pad50Piece
22Computer1Unitself-provide

China Chemiluminescence Nitrogen Analyzer for Solid Liquid Gas Samples 0.1-10000mg/L supplier

Chemiluminescence Nitrogen Analyzer for Solid Liquid Gas Samples 0.1-10000mg/L