Measurement range 0-10% Activated carbon penetration sulfur capacity tester 450ml electrolytic cell

Brand Name:Shandong Shengtai Instrument
Certification:GB/T7702.14-2008
Model Number:ST-64B
Minimum Order Quantity:1
Delivery Time:7-15 work days
Payment Terms:T/T
Contact Now

Add to Cart

Verified Supplier
Location: Jinan Shandong China
Address: Times Headquarters Base, Tianqiao District, Jinan City
Supplier`s last login times: within 13 hours
Product Details Company Profile
Product Details

ST-64B Activated carbon penetration sulfur capacity tester


ST-64B activated carbon penetration capacity testing device is developed in accordance with the national standard GB/T7702.14-2008 "Test methods for coal based granular activated carbon - Determination of sulfur capacity". It is suitable for testing the penetration capacity of coal based granular activated carbon, as well as for testing the penetration capacity of other types of activated carbon.

Experimental principle:

The test sample is burned in a purified air flow at a high temperature of 1050 ℃, and various forms of sulfur in the coal are decomposed and carried to the electrolysis cell by the air flow to combine with water to generate H2SO3. Due to its disruption of the dynamic equilibrium of the original iodine iodine ion pairs in the electrolysis cell, the instrument immediately outputs a current to electrolyze potassium iodide solution to generate iodine, in order to restore the original dynamic equilibrium, which is the Coulomb titration in GB/T214-1996. The amount of current consumed to restore the original dynamic equilibrium is related to the sulfur decomposed during combustion in the laboratory sample, which can be measured and calculated by a microprocessor

Technical parameters

Measurement range of total sulfur:0-10%

The sample combustion analysis time is 3-8 minutes, and the titration endpoint is automatically determined and returned.

Temperature control:500-1150 degrees Celsius, controllable ± 3 ℃. Temperature measurement: accuracy level 0.5; The heating element is a silicon carbon tube, with a high-temperature zone length of ≥ 90 millimeters. The experimental temperature is 1050 ℃± 5 ℃ (the temperature can be adjusted as needed).

Heating rate:25-30 ℃/min, up to 1050 ℃ in about 30 minutes.

Electrolytic cell:with a volume of 450 milliliters and a platinum metal electrode.

The stability of the instrument:It enters a stable state within 1 minute of starting up.

Rated power supply:220V±10% 50HZ

Measurement range of total sulfur0-10%
Combustion analysis time3-8 minutes
Temperature control500-1150℃
Controllable±3℃
Temperature measurement: accuracylevel 0.5
Heating elementSilicon carbon tube
Length of high temperature zone≥90 millimeters
Experimental temperature1050℃±5℃(temperature can be adjusted as needed)
Heating speed25-30℃/min,About 30 minutes to reach 1050 ℃
Electrolytic cellwith a volume of 450 milliliters and a platinum metal electrode.
The stability of the instrumentIt enters a stable state within 1 minute of starting up.
Rated power supply220V±10% 50HZ

ST-64B activated carbon penetration capacity testing device is developed in accordance with the national standard GB/T7702.14-2008 "Test methods for coal based granular activated carbon - Determination of sulfur capacity". It is suitable for testing the penetration capacity of coal based granular activated carbon, as well as for testing the penetration capacity of other types of activated carbon.

Experimental principle:

The test sample is burned in a purified air flow at a high temperature of 1050 ℃, and various forms of sulfur in the coal are decomposed and carried to the electrolysis cell by the air flow to combine with water to generate H2SO3. Due to its disruption of the dynamic equilibrium of the original iodine iodine ion pairs in the electrolysis cell, the instrument immediately outputs a current to electrolyze potassium iodide solution to generate iodine, in order to restore the original dynamic equilibrium, which is the Coulomb titration in GB/T214-1996. The amount of current consumed to restore the original dynamic equilibrium is related to the sulfur decomposed during combustion in the laboratory sample, which can be measured and calculated by a microprocessor

China Measurement range 0-10% Activated carbon penetration sulfur capacity tester 450ml electrolytic cell supplier

Measurement range 0-10% Activated carbon penetration sulfur capacity tester 450ml electrolytic cell

Inquiry Cart 0