1300 Degree Open Ended Thermocouple Furnace

Brand Name:PANRAN
Certification:CE,ISO9001
Model Number:PR320
Minimum Order Quantity:1 pcs
Delivery Time:20~30 working days
Payment Terms:TT, LC, Visa, Western Union, Alipay, Trade assurance, E-checking
Contact Now

Add to Cart

Active Member
Location: Changsha Hunan China
Address: Room 504,Floor 5th,Chuangye Building, Hu-Da park,No186,Guyuan Road,High-Tech District,Changsha,Hunan ,China. (Zipcode: 410205)
Supplier`s last login times: within 48 hours
Product Details Company Profile
Product Details

300~1300℃ Temperature Range Horizontal Design Open-Ended Tube Thermocouple Calibration Furnace


Flexible configuration allows you to calibrate a large variety of probe lengths and diameters
Calibrate thermocouples over a wide temperature range — even high-temperature probes
Prepare comparison calibrations — the ceramic isothermal block holds up to four probes at a time
Alumina ceramic well and isothermal block reduce thermoc


  • Products model selection table
No.NameModelTemperature rangeFurnace sizeDimension (mm)Net Weight (kg)Power (KW)Isothermal block
1Thermocouple calibration furnacePR320A300~1200℃Φ40*600700*370*45026.12.5optional
2Base metal thermocouple calibration furnacePR320B300~1200℃Φ60*60031.52.5/
3Sheathed thermocouple calibration furnacePR320C300~1200℃Φ40*60027.32.5PR1142A
4Thermocouple calibration furnacePR320D300~1300℃Φ40*60026.12.5optional
5Base metal thermocouple calibration furnacePR320E300~1200℃Φ40*60027.32.5PR1145A

  • Detailed technical parameters of the products

1. PR320A/D Thermocouple Calibration Furnace

ModelPR320APR320D
Application

S,R standard thermocouple,working S,R precious and base-metal thermocouple calibration

(need to add PR1145A Isothermal block)

Temperature range300~1200℃300~1300℃
Furnace SizeΦ40mm*600mm
Temperature field centerDeviate from the geometric center does not exceed 10mm
Temperature field widthThe temperature difference within 80mm ≤1℃
Temperature field gradientAdd a coaxial cleaning porcelain tube of about 20mm shall be installed in the furnace,Not exceed 0.4℃/CM within ±20MM range of temperature field center

2. PR320B Base metal thermocouple calibration furnace


Applicationbase-metal thermocouple calibration
Temperature range300~1200℃
Furnace SizeΦ60mm*600mm
Temperature field centerDeviate from the geometric center does not exceed 10mm
Temperature field widthThe temperature difference within 60mm ≤1℃

3. PR320C Sheathed Thermocouples Calibration Furnace

Temperature range300~1200℃
Furnace SizeΦ40mm*600mm
Size of Isothermal blockSize Φ8mm*7pcs, depth 90mm
Temperature field indicators 1 (with Isothermal block)From the bottom of Isothermal hole, the internal temperature difference is ≤0.5 ℃ in the axial direction of 30mm, and the absolute value of the temperature difference between any holes in the same section is ≤0.25 ℃
Temperature field indicators 2 (without Isothermal block)The deviation between the center of the maximum uniform temperature field and the geometric center of the furnace along the axis is ≤10mm;Within the range of a uniform temperature field with a length of ≥60mm and a radius of 14mm, the temperature field between any two points is ≤1℃
Temperature field indicators 3 (with Isothermal block, use a coaxial cleaning porcelain tube of about 20mm)The maximum temperature point in the furnace is ≤20mm away from the geometric center of the furnace, and the maximum temperature ±20mm has a uniform temperature field with the temperature change gradient ≤0.4℃/10mm
Implementation standard
,
Standard itemPR1142A Sheathed thermocouple calibration furnace Isothermal block
Optional itemCoaxial cleaning porcelain tube of Φ20mm

4. PR320E Thermocouple Calibration Furnace

Temperature range300~1200℃
Furnace SizeΦ40mm*600mm
Size of Isothermal blockExternal Φ38mm, innerΦ28mm, depth 150mm
Temperature field indicators 1 (with Isothermal block)Within the axial direction of the effective working area, the absolute value of the temperature difference between any two points is ≤0.5℃;Within the radial radius of ≥14mm, the absolute value of temperature difference between any two points on the same section is ≤0.25℃
Temperature field indicators 2 (without Isothermal block)The deviation between the center of the maximum uniform temperature field and the geometric center of the furnace along the axis is not exceed 10mm;Within the range of a uniform temperature field with a length of ≥60mm and a radius of 14mm, the temperature field between any two points is no more than 1℃
Temperature field indicators 3 (with Isothermal block, but add a coaxial cleaning porcelain tube of about 20mm)The maximum temperature point in the furnace is ≤20mm away from the geometric center of the furnace, and the maximum temperature ±20mm has a uniform temperature field with the temperature change gradient ≤0.4℃/10mm
Implementation standard
,
Standard itemPR1145A Sheathed thermocouple calibration furnace Isothermal block
Optional itemCoaxial cleaning porcelain tube of Φ20mm
China 1300 Degree Open Ended Thermocouple Furnace supplier

1300 Degree Open Ended Thermocouple Furnace

Inquiry Cart 0