OVERVIEW

This product meets the requirements of the Pharmacopoeia issued in 2020,implementing instrument authority management, audit tracking, and electronic signature.It can also be equipped with an exclusive database expert management system.



FEATURES & BENEFITS

  • High intelligence: the test process is fully automatically controlled, without manual intervention;
  • High reliability: high detection accuracy and good reproducibility;
  • High user-friendly: touch screen input, friendly operation interface;
  • Multi-function: can be used to test single-core filters, multi-core filters, and multiple test functions;
  • High integration: the instrument adopts a portable design, small size, and easy to use;
  • High-tech: uses ARM11 controller and Windows CE 6.0 system;
  • High practicality: data storage and real-time printing of test results, in line with GMP requirements.

APPLICATION

  • Disc membrane: various membranes from Φ25mm to Φ300mm
  • Standard cartridge: 2.5″ to 40″, 1 to 9 cores
  • Capsule
  • Mini cartridge
  • Air filter testing 2.5″ to 40″ other parameter cartridges

PRODUCT PARAMENTERS

Power supply/power170-240V AC, 50/60Hz 100W
Maximum operating pressure6000 mbar
Minimum inlet pressure200 mbar
Dimensions40cm(width) x30cm(depth) x 25cm(height)
Test range

forward flow: 0.1-200ml/min

water immersion: 0.05-50ml/min

Bubble point: 700-6000mbar

Test accuracy±3% or ±5‰F.S, whichever is greater
Sensitivity

forward flow: 0.1ml/min

water immersion: 0.01ml/min

bubble point: 50mbar

Repeatability error≤3%
Operating conditions

Ambient temperature: 0℃- 50℃

Relative humidity: 10-80%

Test time

Basic bubble point

Diffusion flow test: 2 min±2min test 7±2min

Water immersion flow test 7±2min

System air tightness: 3±1min

China Zetron FIT-110 Filter Integrity Tester with Automatic Control and Touch Screen Operation supplier

Zetron FIT-110 Filter Integrity Tester with Automatic Control and Touch Screen Operation