Hydrogen Power Multi-Range Handheld Gas Leak Detector With Suction Pump Led Display

Brand Name:UST
Certification:UL 913,CSA, SIRA05 ATEX II 2G EEx ia d IIB T3 IECEx Ex ia d IIB T3
Model Number:HydrogenPower
Minimum Order Quantity:1
Payment Terms:L/C, D/A, D/P, D/P, T/T, Western Union, MoneyGram,Paypal
Place of Origin:Germany
Contact Now

Add to Cart

Verified Supplier
Location: Beijing Beijing
Address: Room 1302,Yufei Building, 42 Dongzhimenwai Street, Dongcheng District, Postcode:100027 Beijing,China
Supplier`s last login times: within 48 hours
Product Details Company Profile
Product Details


Handheld Multi-range Gas Leak Detector for Hydrogen (H2)


OVERVIEW


SPECIFICATION

Measurement range (H2)1 ppm ... 999 ppm, 0,1 Vol%... 4,0 Vol%
Resolution1 ppm / 0,1 Vol% depending on the display range
Response time2s
Time to operation readiness120s
Gas sensorMOX gas sensor element (GGS 1000 + 6000 series)
Flow rate of suction pump40 ml/min
Dimensions (Length x Width x Height)180 mm x 50 mm x 28 mm (without sensor extension flex)
Sensor extension flex (Length)300 mm
Net weight440 g (without charger)
Rechargeable battery power pack

4 x 1,2 V NiMH, to be charged from the mains using the charger supplied with

the gas leak detector

Operating time (power pack is fully charged)6 h
Allowable operating temperature-5 °C… +40°C
Allowable storage and transportation temperature/humidity-25 °C … +70°C / 20… 80 % r.H. (non-condensing)
Allowable operating, transport and storage conditionsAny contamination of the gas sensor must be avoided. The application, transport and storage environment has to be free of any contamination, particularly protected against chemical substances, e.g. silicones. In particular directly contact with substances containing, silicones, sulphurous substances or non-desorbing inorganic components or contaminations (e.g. smoke, fumes, oils, greases or evaporating liquids) may cause damaging the sensor or to changes in the sensor resistance and/or in the sensor characteristics. Possible consequences are reduced sensitivity, display of misleading concentration values, or display of a background concentration.
Conformity2011/65/EU: Restriction of the use of Hazardous Substances Directive (RoHS)

FEATURES & BENEFITS

1. Handheld multi-range gas leak detector with suction pump for the selective detection of Hydrogen (H2)
2. Display of the detected gas concentration value from low ppm range up to "Lower Explosive Limit" (LEL) in Vol%
3. Additional display of the detected gas concentration value from low ppm range up to 1000 ppm by bar graph
4. Automatic measuring range change-over
5. Additional display of concentration thresholds (10, 100, 1000 ppm and 1 Vol%) by LED (optional customized factory calibration)
6. Intermittent tone signal reflecting concentration value by speed and continuous tone signal at 1 Vol% (can be switched off)
7. Automatic zero setting on air after switch on
8. Automatic sensor check with error recognition

China Hydrogen Power Multi-Range Handheld Gas Leak Detector With Suction Pump Led Display supplier

Hydrogen Power Multi-Range Handheld Gas Leak Detector With Suction Pump Led Display

Inquiry Cart 0