Repair 4D V6-2 IU22/IE33 Cable/Connecting Cable Leakage/Damaged Cable


Cable leakage failure reason and solution


Ultrasound probe cable is adopted by high quality of high multi-fiber shielded cable, and good workmanship wires, many hundreds of root, like hair thickness. Due to the bending and torsion of the cable, it makes outside of reserve cable insulating layer will be broken, fracture, revealing the inside line, shielding layer damage, break, corrugated and defect image can be interferenced.

Solution

Due to probe affected by the external force or wrong operators, the probe was not properly placed within for a long time, cable conductor faults. Open the outer leather, lateral horizontal gashing the total shielding wire , pick out the deep crack line, welding is completed, with transparent tape around one layer, use a multimeter to measure whether it is turned on,this wire shielding will be connected and wrapped the tape, to connect the wire shielding and tie up good adhesive tape. As above, If multiple root fracture. After commissioning, everything is all right to better the outermost shielded wire welding, fixed outer holster with high voltage margin plastic tape. in this way, you can continue to use the probe for a long time.




Other hot models we can repair:


BrandModel
C5-2(HD11)
C8-4V(HD11)
L12-3(HD11)
S4-2(HD11)
C5-2(IU22)
L9-3(IU22)
L12-5(IU22)
S3-1(IU22)
S8-3(IU22)
C5-1(CX50/EPIQ/Affiniti)
C10-3V(CX50/EPIQ)
L12-3(CX50/EPIQ/Affiniti)
L12-5(CX50/EPIQ/Affiniti)
S5-1(CX50/EPIQ)
S8-3(CX50/EPIQ)
S12-4(CX50/EPIQ)
C6-2(CX50/Affiniti)
C5-2(ClearVue)
C9-4V(ClearVue)
L12-4(ClearVue)
L12-5(ClearVue)

Common ultrasound Probe Damage(Specialty 3D/4D Volume probes)


Common ultrasound probe damageSolutions
Lens cap damageLens cap replacement
Fluid and oil leaksLens cap replacement, oil reservoir repair and replacement
Cable cutsCable patches, possible cable replacement
Inoperative steeringMotor repair
Connector housing electrical damageMinor electrical repairs, pin module replacement
Weak or dead elements, dropouts in an imageArray ball replacement, oil reservoir repair and replacement

FAQ


How to identify a defective probe?


If the probe has a broken membrane or cracked wire, you can see it immediately. But what if the defect is not noticeable?

Ultrasound probes send and receive ultrasonic signals into the body and create an image. If that image has any dropouts, your probe might have defects.


The most commonly occurring problems are related to wiring sheath or membrane. But you can also get a connector refurbishment, crystal replacement, and more.

Tee transducers are particularly expensive ones and usually get bitten by patients. The tiny defects on tee probes are tough to notice.


You can identify the defects by simple monitoring, such as looking for bite marks, corrosive parts, or crushed tubes.

Even a small defect in the Tee probe could be severe for patients. If at all you have a defective tee probe, PRS can lend you one while repairing yours.