Automatic Control Overpressure Self-Venting Vertical Autoclave
Sterilizer
Vertical Autoclave Sterilizer Features
◆ Hand wheel type door opening structure
◆ Mechanical safety interlocking device
◆ Overpressure self-venting 0.145-0.165Mpa
Model | Material | Volume | Power Supply | Output power | Dimension of sterilizing chamber | Net weight | Package Size |
XFH-30CA | All stainless steel | 30L | 220V | 3.5kW | 280 x 400mm | 47Kg | 510x510x 1170mm |
Vertical Autoclave Sterilizer operating specifications
1. Add water
Out disinfection barrels in the container with water to the high
water (water floating ball float), if the water level below the
water level is, the controller refused a job and alarm of water.
2. the stacked
Sort in advance and bandage good disinfection content inside
sterilization tube, bandage should leave gaps between objects, and
vapor permeability, improve the quality of disinfection
sterilization.
3. The vertical autoclave sterilizer seal
Gently close the container cover, then six thumb screw bolt one by
one to vertical and symmetrical turned up to tighten. Make the
container lid and container seal, no leakage shall prevail.
4. vertical autoclave sterilizer heating
Insert the power cord into the power matching the seat, and there
he put the vent valve handle deflated position, open the control
switch. The controller after the self-inspection, digital tube
display the actual temperature inside.
(1) press the set key, the digital tube display temperature code
back and forth (SP) and temperature, point press to
increase/decrease, needed to adjust to the sterilization
temperature. (initial temperature for the last time)
(2) Press the set key twice, digital tube display time back and
forth code (ST) and time values, click the increase/decrease key,
adjust the sterilization to need time. ST value to "0", the
controller to cancel the function (starting time for the last
time).
(3) press set again, confirm the set parameters, digital tube to
return to display the main interface
(4) Press the start/stop key, sterilizer start work, running lights
flashing, purge valve blocks should be deflated location at this
time. Makes the air inside the sterilizer escape, for valve has a
nasty steam jet, temperature (100 degrees), manually close the
purge valve until the end of the work, sound and light alarm
prompt, work is done.
(5) the work if you want to cancel the sterilization command, press
the start/stop button 3 seconds, the controller to immediately stop
work, buzzer alarm prompt, press any key noise elimination.
(6) Water shortage in the process of work, the controller
automatically stop working, and alarm prompt lack of water. Press
any key noise elimination. (if need to continue to work, should be
rated pressure gauge to zero up water)
5. cooling
When after the completion of vertical autoclave sterilization
should be shut off the power, after waiting for its cooling to the
gauge pressure to zero, then open valve or the container cover.Do
not completed in sterilization, immediately open valve or container
cover, solutions in order to avoid boiling overflow, cause an
accident.