PHiliph VS2 LCD Display Used
PHiliph SureSigns VS2 Vital Signs Patient Monitor LCD Display with Flex Cable Used-Good Medical Device Spare Parts
► Brief Description
PHiliph VS2 Patient Monitor LCD Display - is a very important part of the VS2 patient monitor. It can clearly display all monitored patient physiological characteristic parameters and present patient data to medical staff for reference, facilitating clinical judgment and decision-making.
► Specification
Product Category | Monitoring |
Product Type | Spare Parts |
Product Item | Display |
Display Size | 4.3" |
Display Type | WVGA |
Display Resolution | 480 pixels x 272 lines |
Machine Application | VS2 |
Condition | Used - Good |
Warranty | 90 Days |
Delivery Date | 3 - 5 Working Days |
Payment Method | Western Union, Bank T/T, Paypal |
Transportation | DHL, FedEx, UPS, TNT, EMS, By Air, etc. |
Material | Silicon, Plastic, Metal, Glass, etc. |
► Detailed Photos
► Test
When using it, you can test the display first. Here are the steps:
1. In the "System Diagnostics" window, rotate the wheel to highlight the "Display Test" button, and then press the wheel to start the test.
The display test draws a series of patterns on the monitor screen. Each pattern is displayed for five seconds. After the last pattern is drawn, it will restart from the first pattern.
2. To exit the test, press or turn the wheel.
► Display Problems
During the use of SureSigns VS2 Patient Monitor Display, some faults may occur. Here are some possible faults and causes, as well as corresponding solutions:
llFault: The power is on, but the monitor screen is blank.
1. Possible cause: The monitor is in power saving mode.
Solution: Press the power button to end power saving mode.
2. Possible cause: The mainboard is faulty.
Solution: Replace the mainboard.
3. Possible cause: The LCD is damaged.
Solution: Replace the LCD.
llFault: The monitor displays random/distorted graphics with a white background.
1. Possible cause: The LCD cable is not connected correctly.
Solution: Turn on the monitor and reinsert the LCD signal cable.
2. Possible cause: The LCD is damaged.
Solution: Replace the LCD.
3. Possible cause: The mainboard is faulty.
Solution: Replace the mainboard.
llFault: After turning on the monitor and the startup screen or color bars, the monitor is stuck.
Possible cause: The mainboard is defective.
Solution: Replace the main PCB.