PHiliph SureSigns VS2 Vital Signs Patient Monitor LCD Display with Flex Cable Used-Good Medical Device Spare Parts

Brand Name:PHiliph
Model Number:VS2
Minimum Order Quantity:1 Piece
Delivery Time:3-5 working days
Payment Terms:T/T, Western Union, Paypal
Price:USD
Contact Now

Add to Cart

Verified Supplier
Location: Guangzhou Guangdong China
Address: TianHe District,GuangZhou,GuangDong
Supplier`s last login times: within 33 hours
Product Details Company Profile
Product Details

PHiliph VS2 LCD Display Used

PHiliph SureSigns VS2 Vital Signs Patient Monitor LCD Display with Flex Cable Used-Good Medical Device Spare Parts


Brief Description

PHiliph VS2 Patient Monitor LCD Display - is a very important part of the VS2 patient monitor. It can clearly display all monitored patient physiological characteristic parameters and present patient data to medical staff for reference, facilitating clinical judgment and decision-making.


Specification

Product CategoryMonitoring
Product TypeSpare Parts
Product ItemDisplay
Display Size4.3"
Display TypeWVGA
Display Resolution480 pixels x 272 lines
Machine ApplicationVS2
ConditionUsed - Good
Warranty90 Days
Delivery Date3 - 5 Working Days
Payment MethodWestern Union, Bank T/T, Paypal
TransportationDHL, FedEx, UPS, TNT, EMS, By Air, etc.
MaterialSilicon, Plastic, Metal, Glass, etc.

► Detailed Photos


► Test

When using it, you can test the display first. Here are the steps:

1. In the "System Diagnostics" window, rotate the wheel to highlight the "Display Test" button, and then press the wheel to start the test.


The display test draws a series of patterns on the monitor screen. Each pattern is displayed for five seconds. After the last pattern is drawn, it will restart from the first pattern.


2. To exit the test, press or turn the wheel.

► Display Problems

During the use of SureSigns VS2 Patient Monitor Display, some faults may occur. Here are some possible faults and causes, as well as corresponding solutions:

llFault: The power is on, but the monitor screen is blank.

1. Possible cause: The monitor is in power saving mode.

Solution: Press the power button to end power saving mode.

2. Possible cause: The mainboard is faulty.

Solution: Replace the mainboard.

3. Possible cause: The LCD is damaged.

Solution: Replace the LCD.

llFault: The monitor displays random/distorted graphics with a white background.

1. Possible cause: The LCD cable is not connected correctly.

Solution: Turn on the monitor and reinsert the LCD signal cable.

2. Possible cause: The LCD is damaged.

Solution: Replace the LCD.

3. Possible cause: The mainboard is faulty.

Solution: Replace the mainboard.

llFault: After turning on the monitor and the startup screen or color bars, the monitor is stuck.

Possible cause: The mainboard is defective.

Solution: Replace the main PCB.


Feel free to contact us for purchase:

My WhatsApp/WeChat/phone: +86 151 8004 7496

My E-mail: dhy.jasmine@gmail.com



Thanks for your interest in PHiliph SureSigns VS2 Vital Signs Patient Monitor LCD Display with Flex Cable Used-Good Medical Device Spare Parts, please feel free to contact me. Thank you for your time and wish you have a happy life.

China PHiliph SureSigns VS2 Vital Signs Patient Monitor LCD Display with Flex Cable Used-Good Medical Device Spare Parts supplier

PHiliph SureSigns VS2 Vital Signs Patient Monitor LCD Display with Flex Cable Used-Good Medical Device Spare Parts

Inquiry Cart 0