Product Describe

This is a USB interface board for EDAN F6 fetal monitor. As its name suggests, it is an interface for connecting external USB. It can transmit the data of the device to the external device, such as sending the collected fetal heart rate curve, uterine contraction pressure, uterine contraction record and other data to a USB flash drive, or transmitting to a printer to print a paper report. It can also receive external signals, such as the PC upgrade program required for system upgrade, the signal transmitted by the external keyboard, etc. It is the key hub for the digital upgrade of Edan F6.

It can mainly reflect the function in some clinical scenarios; emergency printing in the delivery room: insert a USB flash drive to back up important data (fetal heart rate, uterine contraction pressure, etc.) to prevent data loss caused by equipment failure. It can also move data in an emergency (shortage of thermal paper), find other solutions, and have multiple options to avoid delaying clinical diagnosis and treatment.

Data synchronization in multiple hospital areas: export data to a mobile USB flash drive through the USB interface board, share monitoring records in multiple hospital areas, discuss medical diagnosis plans across hospital areas, reduce the risk of high-risk pregnancy, and improve medical efficiency. Centralized equipment maintenance: To upgrade the monitor firmware in batches, you need to use the PC upgrade program in conjunction with the USB Hub.


Product Picture


Functions

1. Manage medical data

Encrypted export data: The device supports encrypted storage of fetal monitoring data to protect patient privacy and prevent data leakage.

Automatic naming: Generate files by "date + time + pregnant woman ID" (such as 20240815_1430_510123.pdf), which is convenient for medical staff to find relevant data.

2. Equipment system maintenance

System upgrade: The accessory comes with a Micro-B interface, which can be connected to a computer and the device can be upgraded through the official program.

Log extraction: Insert a USB flash drive to automatically export the device operation log (.edl format), which is convenient for professional maintenance personnel to analyze faults based on the log data.

3. Extended application

Wireless printing: The USB interface can be connected to an external WiFi adapter to connect to a wireless printer to print the required files.

Generate electronic medical records: An external barcode scanner can quickly enter the pregnant woman's hospitalization number and synchronize the hospital medical record management system.


Products Parameters

Interface type

USB 2.0 Host ×2 (Type-A female port)

USB 2.0 Device ×1 (Micro-B type)

Working environment


Temperature:0℃~40℃

Humidity


20%~80% RH (non-condensing)
Data transmission rateHigh-speed mode (480Mbps), measured fetal monitoring data transmission speed ≥15MB/min

Protocol support


Storage protocol: USB Mass Storage Class (U disk direct reading)

Printing protocol: ESC/POS medical simplified version

SizeSmall
WeightLight
Delivery3-5 working days
Warranty3 months

Installation and troubleshooting

1. USB interface board replacement steps

① First check whether the power is cut off, then open the back cover of the host;

② Remove the interface board fixing screws and pull out the old interface board;

③ Install the new interface board according to the meaning of the interface and tighten the screws;

④ Turn on the computer to check whether the USB interface function is normal: Enter [System Settings] → [USB Function Self-Test].

2. Troubleshooting

USB Host overcurrent--Check whether the external device is compatible

File system incompatible--Format the USB disk

Abnormal isolation impedance--Return to the factory for inspection