Carefusion Bird Flow Sensor 15972 New Original For VELA Ventilator Introduction


The Carefusion Bird 15972 Flow Sensor is a new, original component designed for use with VELA ventilators. This flow sensor is essential for


accurately measuring and monitoring gas flow in ventilation systems, ensuring optimal performance and patient safety.


The Carefusion Bird 15972 Flow Sensor is a reliable, high-quality component that is easy to install and compatible with VELA ventilators.


Designed for long life and accurate measurements, it is an essential component for medical staff and patients requiring ventilatory support.


When used with VELA ventilators, the Carefusion Bird 15972 Flow Sensor allows healthcare professionals to rely on its accuracy and


performance to provide effective respiratory support to patients in need. Trust the Carefusion brand when it comes to quality and reliability of


ventilation components.


Specification of Carefusion Bird Flow Sensor 15972 New Original For VELA Ventilator


Product NameCarefusion Bird Flow Sensor 15972 New Original For VELA Ventilator
BrandVela
Part No15972
MaterialPlastic
ColorYellow
MOQOne piece
Warranty90 days

Oxygen Sensor part number 15972 Disassembly and Instalation


For older Vela units, perform the following steps:


Disassembly


1. Disconnect the power cord, top cover, battery tray, left side panel, and bottom panel.


2. Lay the unit horizontally.


3. Remove the 1-inch mounting nut from the manifold base and remove the sensor from the manifold.


NOTE

Remove the O2 sensor only if complete removal is not required.


Installation


1. Ensure the O-ring is installed on the oxygen sensor.

2. Lay the unit on its side, place the sensor on the manifold, and install the 1-inch mounting nut.

3. Install the bottom panel, left side panel, battery tray, top cover, and power cable.


For newer Vela units (typically 1400 series and above), perform the following steps:


Disassembly


Remove the screws on the EMI shield containing the foam ambient air filter located on the rear lower portion of the Vela and lift the shield and filter together. . Pull the O2 sensor firmly back to remove.


Installation

Install and replace the new O2 filter. Replace the EMI shield with the foam surround air filter and retighten the screws holding it in place.