Calibrating the CO2 Module


Important - Since the display uses the current date and time as the calibration date/time, ensure that the system date and time are set correctly.

NOTE - The monitor always shows CO2 module operating time (operating time) in CO2 test window.

1. Connect the other end of the FilterLine hose (if not already connected) to the monitor's CO2 input; leave the other end unconnected.

2. In the CO2 test window, highlight the scroll wheel Start calibration and press the scroll wheel. A confirmation window will appear prompting you to connect the gas to the CO2 module.

3. Connect the other end of the FilterLine hose to the inlet of the 5% gas concentration gas regulator.

4. Open the gas control valve to allow 5% CO2 gas to flow through the monitor.

5. In the confirmation window, turn the dial to the Yes position and press the dial to start the calibration.

The display will start calibrating the CO2 module and display the calibration value when it is complete.

NOTE—The calibration process can take up to a minute.

6. Turn off the airflow and turn the wheel to return to the CO2 test window. End the test by pressing the selector.

7. If the calibration is successful, the date and time of the last calibration will be reset in the CO2 test window. Check this by opening the CO2 test window again.

8. If the module cannot be calibrated, an error message will appear. CO2 module replacement


Specification of PHiliphh VM8 Co2 module 463664015441 .

Product NamePHiliphh VM8 Co2 module 463664015441
BrandPHiliphh
Part No463664015441
Use forPHiliphh VM8 patient monitor
ConditionUsed & New
StockSufficient

Removing the SureSigns VM8 CO2 Module

To remove the CO2 module:

1. Disconnect the power cord behind the monitor.

2. Remove the battery from the display.

3. Remove the four Torx T10 screws and washers (3 on the bottom and one on the top) from the CO2 module cover.

4. Remove the CO2 module from the rear of the monitor housing. And disconnect the CO2 wire that connects to the CO2 board connector.

FAQ:

  • what’s your MOQ

Our MOQ is one piece .Welcome to take samples to test our quality .


  • When can i receive my orders ?

Usually it will take around 2-3 working days to prepare for the order after payment and you will receive it around 5-7 working days .


  • Do you have product list ?

As our stock is changing ,right now we don’t have products list .If you would like to know what we have ,feel free to check our website ,and send us pictures to check price

  • Can i have a test video before placing order?

Sure .Video can be sent as well before the shipment .


If you're interested in 463664015441 PHiliphh VM8 Patient Monitor Spare Part Co2 Module ,feel free to contact us .Thank you .


China 463664015441 PHiliph VM8 Patient Monitor Spare Parts Co2 Module supplier

463664015441 PHiliph VM8 Patient Monitor Spare Parts Co2 Module

Inquiry Cart 0