6802-30-66765 Mindray T5 Patient Monitor Parts Integral Module Frame Rack - patients-monitor

6802-30-66765 Mindray T5 Patient Monitor Parts Integral Module Frame Rack

Brand Name:PHiliph
Model Number:6802-30-66765
Minimum Order Quantity:1pcs
Delivery Time:3-5 working days
Payment Terms:T/T(wire transfer), Western Union, Paypal
Packaging Details:Carton
Contact Now

Add to Cart

Verified Supplier
Location: Guangzhou Guangdong China
Address: TianHe District,GuangZhou,GuangDong
Supplier`s last login times: within 33 hours
Product Details Company Profile
Product Details

6802-30-66765 Mindray T5 Patient Monitor Integral Module Frame Rack


Integral Module Rack


The patient monitors are available in two integrated modular racks: 2-slot and 5-slot. The control board contains a NIOS II FPGA. Realizes protocol conversion and infrared communication between host and parameter module

The module rack communication card can have 2 or 3 slots. The 3-slot communication card communicates directly with the motherboard. The 2-slot communication card is connected and controlled by the 3-slot communication card. The 3-slot communication card has a communication control function. The 2-slot communication card consists of an infrared circuit and a module power supply circuit. The RS422 driver board is located on the 3-slot communication board.


Removing the Integral Module Rack


  • Disconnect the cable that connects the integrated module frame to the motherboard. Unscrew the five M3 × 8 screws and remove the built-in module holder.

  • Disconnect the cable that connects the communication board of the MPM frame to the communication board of the 3-slot module housing. Release the two buttons and remove the NiosII module from the 3-slot module holder.

  • Loosen the eight M3 × 6 screws, and then remove the 3-slot modular rack communication card and the MPM modular rack communication card.

Note: Do not forget the ground spring when assembling.

  • Loosen the hex nut with the sleeve. Then separate the disc, spring and contact pin.


Module Rack Failures


SymptomsPossible CauseCorrective Action
Integral module rack cannot identify parameter modulesModule failureReplace the parameter module. If a new module is identified, the original module is defective.
Cable defective or poorly connected

1. Check that the cables from the 3-slot module rack communication board to the MPM module rack communication board and from the module rack to the main board are properly connected.

2. Check that the connection cables and plugs are not damaged

Wrong communication board software revisionUpgrade the program of the module or Integral module rack.
Module (in some slots) unrecognizedReplace the corresponding module rack communication board.
Power supply to integral module rack abnormal1. Check if voltage between two contact screws in any slot reaches 12VDC. If yes and the parameter module functions, PCB assembly in the SMR might fail. 2. If there is no 12V sent to the integrated module rack, check that power module output voltage to mother board reaches 12V DC. If yes, mother board might fail.
3-slot or MPM module rack communication board failureReplace the 3-slot or MPM module rack communication board.
Nios II module failureReplace the Nios II module.