Product Details
6802-30-66765 Mindray T5 Patient Monitor Integral Module Frame Rack
Integral Module Rack
The patient monitors are available in two integrated modular racks:
2-slot and 5-slot. The control board contains a NIOS II FPGA.
Realizes protocol conversion and infrared communication between
host and parameter module
The module rack communication card can have 2 or 3 slots. The
3-slot communication card communicates directly with the
motherboard. The 2-slot communication card is connected and
controlled by the 3-slot communication card. The 3-slot
communication card has a communication control function. The 2-slot
communication card consists of an infrared circuit and a module
power supply circuit. The RS422 driver board is located on the
3-slot communication board.
Removing the Integral Module Rack
- Disconnect the cable that connects the integrated module frame to
the motherboard. Unscrew the five M3 × 8 screws and remove the
built-in module holder.
- Disconnect the cable that connects the communication board of the
MPM frame to the communication board of the 3-slot module housing.
Release the two buttons and remove the NiosII module from the
3-slot module holder.
- Loosen the eight M3 × 6 screws, and then remove the 3-slot modular
rack communication card and the MPM modular rack communication
card.
Note: Do not forget the ground spring when assembling.
- Loosen the hex nut with the sleeve. Then separate the disc, spring
and contact pin.
Module Rack Failures
Symptoms | Possible Cause | Corrective Action |
Integral module rack cannot identify parameter modules | Module failure | Replace the parameter module. If a new module is identified, the
original module is defective. |
Cable defective or poorly connected | 1. Check that the cables from the 3-slot module rack communication
board to the MPM module rack communication board and from the
module rack to the main board are properly connected. 2. Check that the connection cables and plugs are not damaged |
Wrong communication board software revision | Upgrade the program of the module or Integral module rack. |
Module (in some slots) unrecognized | Replace the corresponding module rack communication board. |
Power supply to integral module rack abnormal | 1. Check if voltage between two contact screws in any slot reaches
12VDC. If yes and the parameter module functions, PCB assembly in
the SMR might fail. 2. If there is no 12V sent to the integrated
module rack, check that power module output voltage to mother board
reaches 12V DC. If yes, mother board might fail. |
3-slot or MPM module rack communication board failure | Replace the 3-slot or MPM module rack communication board. |
Nios II module failure | Replace the Nios II module. |
Mother board failure | Replace the mother board. |
If you're interested in 6802-30-66765 Mindray T5 Patient Monitor Integral Module Frame Rack ,feel free to contact us .Thank you .
Company Profile
China World Technology Medical Equipment Service Group is a company
engaged in medical equipments/repair parts and accessories products
trading company. The company was established in September 2005, has
many famous enterprises at home and abroad (such as America USCO
company) to establish a long-term friendly business relationship.
Company mainly engaged in patient monitor and other medical machine
products. Since its inception continue to absorb a variety of
talents, improve product quality, improving staff quality, in order
to grow their companies. Strict quality guarantee system and
perfect management system, high-quality products after-sales
service is our foothold in this, my company of "quality first,
reputation first" principle, provide customers with quality and
quantity of various types of products. Always uPHiliphold the
"quality, integrity and pragmatic, motivated, service-oriented"
business PHiliphilosoPHiliphy, and applies the company's management
and operating among. The face of fierce competition, the company
system is constantly being improved, relying on science and
technology, continuously improve the technology content of products
sold, for society, customers and companies to create a higher
market value. Our company has been in good faith to create
enterprises, to credit management market, has won a good
reputation, also won the respect of their domestic counterparts.
Business strictly in accordance with relevant state laws,
regulations and rules of the WTO requirements management according
to law, actively participate in regional economic.Business
prospects and good, the next time, our company will expand the
scale of operation and steady development of corporate economic,
sincerely seeking partners, good faith cooperation and seek common
development, writing a new chapter in medical equipment field.