TR60-F Mindray IMEC7 IMEC8 IMEC10 IMEC12 IPM8 IPM9 IPM10 IPM12 MEC T5 T6 T8 Series Patient Monitor Printer Recorder


About Recorder

The recorder receives data from the motherboard and sends it to the thermal print head for printing. There is a hard button (start/stop recording) and a green LED on the front of the recorder. It is connected to the motherboard.


The following diagram shows its operating principle:


ModuleDescription
Power interfaceIntroduces a DC from the main board.
Power moduleThe input power is converted to a voltage suitable for each module and then forwarded to each module.
CPUCommunication between control modules.
Signal interfaceControls the communication between the main board and the recorder CPU.
Motor drive circuitReceives control signals from the CPU and forwards them to the stepper engine.
Key&LED interfaceIncludes a button and an LED which are directly controlled by the CPU.

Recorder

TypeThermal array
Speed25 mm/s, 50 mm/s
Trace3


Recorder Check


1. Print an ECG trace. The recorder should print correctly and the printout should be clean.

2. If the configuration recorder runs out of paper and other problems occur, the display should provide the required information. When the problem is resolved, the recorder should work normally.


3. Enable automatic alarm logging for all parameters and set a limit outside the alarm limit set for each parameter. In the event of a parameter alarm, the corresponding alarm recording must be activated.


Recorder Failures


  • No printout

Possible CauseCorrective Action
Recorder module disabled

1. Check if the recorder status indicator is turned on

2. If yes, enable the module in the [Factory Maintenance] menu. If not, look at other possible reasons.

Paper reversedRe-install the paper roll.
Cable defective or poorly connected

1. Check that the cable between the recorder and the motherboard is properly connected.

2. Check the connecting cables and connectors for damage.

Recorder power supply failureCheck if the power module outputs 5 V DC and 12V DC correctly.
Recorder failureReplace the recorder

  • Poor print quality or paper not feeding properly
Possible CauseCorrective Action
The paper roll is not installed correctlyStop the recorder and reinstall the paper roll.
Print head dirty

1. Check the thermal print head and paper roller for foreign objects.

2. Clean the thermal printhead with a suitable cleaning solution.

Print head failureReplace the print head.
Recorder failureReplace recorder.

Removing the Recorder (Optional)


1.Push the two clips aside and take out the recorder driver board, as shown below:

2.Unscrew the two PT2X6 screws and disconnect the cable and flexible cable between the recorder driver board and the recorder keyboard, as shown below. Then remove the thermal print head and recorder driver board.

3.Remove the keyboard by removing the two PT2X6 screws shown below.