Product Details

Nihon Kohden Cardiolife TEC-7621C Defibrillator Blood Pressure Plate UR-0257 6190-022986A S1


Brief Introduction

This Blood Pressure Plate is work for Nihon Kohden Cardiolife TEC-7621C Defibrillator.

CWT Technology HK Limited, located at Guangzhou, China. During these years of marketing and expanding, we have accumulated a great wealth of experience both in product distribution and customer service. We are a trust worthy medical equipment supplier as we have been engaged in this industry for so many years. Customer will received satisfactory goods.

We understand the importance of the concept of win-win cooperation, we commit ourselves to protect the medical equipment market, satisfying all the customers' demands is our permanent pursuit.

We are looking forward to establishing a long cooperative relationship with you.


NIBP Specifications


ParameterSpecification
Display parametersSystolic pressure, mean pressure, diastolic pressure, cuff pressure
Measuring range:0 to 300 mmHg
Accuracy:±3 mmHg (0 mmHg ≤ NIBP < 200 mmHg)
±4 mmHg (200 mmHg ≤ NIBP ≤ 300 mmHg)
Measurement mode:Manual / STAT (continuous) / Automatic (periodic)

Alarm:

Adult

  • Upper limit range: 15 to 260 mmHg in 5 mmHg steps, OFF
  • Lower limit range: OFF, 10 to 255 mmHg in 5 mmHg steps

Neonate

  • Upper limit range: 10 to 125 mmHg in 5 mmHg steps, OFF
  • Lower limit range: OFF, 5 to 120 mmHg in 5 mmHg steps


NIBP Mesurement Troubleshooting


MessagePossibleCause Action
NIBP air hose disconnectedThe air hose is disconnected from the NIBP socket.Connect the air hose properly. When NIBP monitoring is not necessary, press the SILENCE ALARM key to silence the alarm. The NIBP display can be cleared.
The air hose is damaged.Replace the air hose with a new one.
Faulty NIBP unit.Replace the NIBP unit.
NIBP air leakThe cuff pressure does not change after inflation even after a certain period of time.Connect the cuff to the air hose properly.
Connect the air hose to the socket properly.
The cuff or air hose is damaged.Replace the cuff or air hose with a new one.
Faulty NIBP unit.Replace the NIBP unit.
NIBP measurement time outThe measuring time exceeded the specified time due to arrhythmia or noise.If the cause is arrhythmia, measure by invasive blood pressure measurement. Remove the cause if due to noise.
Faulty NIBP unit.Replace the NIBP unit.
Faulty NIBP unit.Replace the NIBP unit.