Mindray IMEC8 Patient Monitor Display 8.4" TM084SDHG01 M1P6563706400


The iMEC8, iMEC10 and iMEC12 support 8", 10" and 12" LCD monitors respectively. The displays are connected to the motherboard via display cable and backlight cable. The connection cables of the three patient monitors are different and unreliable. The signal cable is pretty bad. If there is a problem with the display Consider replacing the display cable first.


Touch screen control panel and touch screen

Fully configurable iMEC patient monitor supports touch screen. The touch screen can be controlled by touch and can be calibrated. The touch screen is connected to the touch screen control panel and the main board. The touch screen control panel operates on the touch screen and allows communication with the patient monitor. The touch screen control panel communicates with the main board through the I2C interface, and the touch screen is connected via FPC.


Specification


Size8.4”
Resolution800 x 600 pixels
Dimensions268 x 210 x 114mm
WaveformUp to 8
External display1 display through VGA

Display Failures


SymptomsPossible CauseCorrective Action

The display is

blank or black.



Cable defective

1. Check whether the cable between the monitor and the motherboard and the backlight cable are connected correctly.

2. Inspect cables and connectors for damage.

Main board defectiveReplace the main board
isplay defectiveReplace the display.

SymptomsPossible CauseCorrective Action

Images

overlapped or

distorted


main board error

Replace the main board, or upgrade the main

board with the upgrade software.


Cable defective

1. Check whether the cable between the monitor and the motherboard and the backlight cable are connected correctly.

2. Inspect cables and connectors for damage.


SymptomsPossible CauseCorrective Action

Snows or flashing

specks on the

display.The display does

not function.


Cable defective

1. Check that the cable between the secondary monitor and the monitor is properly connected.


2. Inspect cables and connectors for damage.

The connector board

failed.


Replace the connector board .
The main board failed.Replace the main board.


SymptomsPossible CauseCorrective Action

Touchscreen does

not respond


Touchscreen disabled

Check if there is a

displayed above

the [

Main Menu

] quickkey. If yes, press and hold

the [

Main Menu

] quickkey for more than 3

seconds to enable the touchscreen


Cable defective

1.Check whether the cables between the touch screen and the touch screen control board, and between the touch screen control board and the main board are connected correctly.

2. Check for loose cables and connectors.

Touchscreen control

board defective


Replace the touchscreen control board
Touchscreen defectiveReplace the touchscreen.
The main board failed.eplace the main board.

Touch position

invalid


Touchscreen not

calibrated


Calibrate the touchscreen.