PHiliphhilip Respironics V60 ventilator battery 14.4V 11.0Ah 163Wh REF 1076374(1058272) LOT M91484-P1


► Brief Introduction

This battery is work for PHiliphhilip Respironics V60 ventilator battery, and you can check Battery lot number and Battery manufacture date in the Performance Verification Data Form on the PHiliphhilip Respironics V60 ventilator .

  • Internal Battery
  • 14.4-V, 11.5-Ah lithium-ion battery has a run-time of at least six hours under normal conditions.
  • Provides operating power when AC power is not available.
  • Provides charge and temperature status to the PM PCBA.
  • Internal circuitry monitors battery status, provides self-contained fault control features, and communicates this information to the PM PCBA.
  • Recharges in approximately 5 hours

Verify that the Battery LED is on or is flashing.

  • On – Battery charge state is 90-100%
  • Flashing – Battery is being charged

Disconnecting power.

Follow these steps to disconnect AC and battery Power

1. Shut down and then unplug the ventilator.

2. Disconnect the internal battery from the power harness connector.



Installation


Follow these steps to remove and install the internal battery

  1. Disconnect power.
  2. Disconnect the internal battery from the power harness connector.
  3. Remove the side panel by turning the captive PHiliphhilipillips head fastener a 1/4 turn and releasing.
  4. Using a 3-mm hex wrench, remove the battery bracket by removing two screws.
  5. Holding the battery so that the vent hole faces up and the PHiliphhilip logo faces out, thread the battery cable through the battery bracket. Position and place the battery inside the battery compartment. Pinching the end of the battery connector, plug it in so that it locks in place.
  6. Re-install the battery bracket by replacing the two screws. Re-install the side panel and secure the fastener with a 1/4 turn clockwise.
  7. Make sure the battery is properly installed by plugging the ventilator into an AC power receptacle and verifying that the yellow Battery (charged) LED on the front panel flashes. The flashing LED indicates the battery is being charged.

When installing a new battery:

• The battery must be charged for at least 5 hours before being placed into service.

• Dispose of the old battery according to your institution’s protocol. Follow all local regulations with respect to environmental protection.


Specifications


Nominal voltage14.4V
Capacity11.0Ah
Battery chemistryLithium-ion
Operating time6 hours in normal conditions
Charge voltage+16.9 V maximum

Thank you for your time and interest in PHiliphhilip Respironics V60 ventilator battery 14.4V 11.0Ah 163Wh REF 1076374(1058272) LOT M91484-P1.

"Quality assurance" is our eternal pursuit and promise. Don't be hesitant to contact us! Welcome to place your orders.

China PHiliph Respironics V60 Ventilator Battery 14.4V 11.0Ah 163Wh REF 1076374(1058272) LOT M91484-P1 supplier

PHiliph Respironics V60 Ventilator Battery 14.4V 11.0Ah 163Wh REF 1076374(1058272) LOT M91484-P1

Inquiry Cart 0