MR340E Fisher& Paykel Humidification Chamber For Infant/Pediatric


MR340 is a reusable humidification chamber for use with Fisher & Paykel heated bases. This chamber is used to ensure the humidification of intubated or non-intubated patients. Its low compressible volume makes the MR340 ideal for infants or newborn applications.


Product Specification

Gas portsInlet15mm Female
Outlet22mm Male
Maximum Operating20kPa
Compliance0.15-0.35mL/cm H2O
Maximum Peak Flow70L/min

With ScrollWithout scroll

Compressible Volume


Full130mL135mL
Empty360mL370mL
Maximum water capacity230mL235mL

Material Used In Construction

BaseAluminium
ScrollAluminium
WickAbsorbent Paper
DomePolysulPHiliphhilipone
GasketSynthetic Rubber

Test Condition

Humidifier :MR730(similar results can be expected using an MR850/MR700/720 )

Humidification chamber :MR340 without scroll

Maximum Heater Plate Setting :110 ℃ using 150w heater plate

Ambient air temperature :22-24℃


With Heater wire

Recommended flowrates (L/min)

1.5 L/mim

(Minimum)

30L/min

(typical )

70L/min

(Maximum)

Pressure Drop(cm H2O)0.010.381.96

Time to Rearch Airway

Temperature =37℃

5 min5 min5 min

Humidity at airway outlet

Set Temp=37℃

44mg/L43.5mg/L42mg/L

Without Heater wire

Recommended flowrates (L/min)

5 L/mim

(Minimum)

30L/min

(typical )

50L/min

(Maximum)

Pressure Drop(cm H2O)0.010.381.02

Time to Rearch Airway

Temperature =37℃

<20 min15 min15 min

Humidity at airway outlet

Set Temp=37℃

44mg/L(37°)43.5mg/L(36°)42mg/L(35°)

Humidifier :MR410

Humidification chamber :MR340 with scroll

Ambient air temperature :22-24℃

Circuit:1.3*15mm Smoothbore Tubing (900MR529)


Recommended flowrates (L/min)

5 L/mim

(Minimum)

20L/min

(typical )

40L/min

(Maximum)

Pressure Drop(cm H2O)0.010.050.18
Time to Reach Operating Temperature25 min30 min40 min

Liquid Output(mg/L)

(Airway Temp Reached ℃)


Power Setting=5

Power setting =9


30mg/L(29.5°)

39mg/L(35°)



43.5mg/L(32.5°)

46mg/L(38°)


42mg/L(30°)

34mg/L(32°)


Warning


  • Do not exceed the maximum fill level. If the chamber is overfilled, fluid can enter the breathing circuit.
  • Do not fill the room with water above 37°C.
  • If using a water supply unit, make sure to place the water bag below the level of the humidifier after filling the water chamber.
  • Make sure to regularly monitor the water level in the chamber. Supplement if necessary.
  • Make sure the humidifier is not tilted. Tilting the humidifier may cause water to enter the breathing circuit.
  • Do not touch the heater plate or the chamber base. Surfaces may exceed 85°C.
  • Do not exceed the maximum peak flow of 70L/min.

Cleaning

The cleaning chamber can be cleaned by any of the following methods


Autoclave: 136°C (277°F) @ 220kPa (32psi) for 4 minutes or 120°C (248°F) @ 96kPa (14.1psi) for 15 minutes

Ethylene oxide: 55°C (131°F)

Pasteurization Chemical Immersion:

Solutions such as Mesoxazine, Pyrrole, Control 3, Solution 2 and Cidex have been tested according to the manufacturer's instructions. Other brand names with similar active ingredients can also be substituted.