Mindray BeneHeart R3 ElectrocardiograPHiliphhilip ECG Machine patient monitor Parameter Board 115-018611-00


►Brief Introduction

This Parameter board is responsible for ECG data acquisition of Mindray BeneHeart R3 ElectrocardiograPHiliphhilip ECG Machine.


Removing the Parameter Board

1. Unscrew the four M3×6 tapping screws.

2. Disconnect the cable between the main control board and the parameter board.


► Enviornment

To avoid explosion hazard, do not use the equipment in the presence of flammable anaesthetics, vapours or liquids. The environment where the equipment will be used should be reasonably free from vibration, dust and corrosive substances. If these conditions are not met, the accuracy of the system may be affected and damage may occur.

Check and make sure that there is no electromagnetic interference source around the equipment, especially large medical electrical equipment such as radiological equipment and magnetic resonance imaging equipment etc. Switch off these devices when necessary.

Keep the examination room warm (no less than 18℃) to avoid muscle action voltages in ECG signal caused by cold.


► ECG performance test and calibration

The ECG waveforms displayed and printed out should be clear and complete without obvious noise. The amplitude of lead II waveform is about 10 mm and diagnosis indicate sinus rhythm and HR 80 bpm.

ECG Lead Off prompt behaves correctly.


► Prompt Message

MessageTriggerAction
ECG data insufficientIn the situation that pre-acquisition is enabled, the equipment has not acquired 10 seconds of ECG data when auto measurement is started.Wait till sufficient data is acquired.
Acquiring...The equipment is acquiring 60-second ECG data when rhythm measurement is started.Wait till 60 seconds of countdown is reached. To stop acquisition, press the [Rhythm] key.

► Problems

ECG data displays unacceptable noise.

Possible Cause:

  1. Patient movement.
  2. Improper filter setting.
  3. Poor skin preparation.
  4. Electrode problem.
  5. Patient cable problem.
  6. Wrong accessories are used or mix electrode types and brands.

Corrective actions

  1. Tell the patient not to move during ECG acquisition.
  2. Check the settings of the filters are appropriate.
  3. Prepare the patient before ECG acquisition.
  4. Verify the electrodes are applied correctly. Check for defective or expired electrodes.
  5. Check for defective, broken or disconnected patient cable.
  6. Use specified accessories. Do not mix electrode types or brands.

►FAQ

Q1

How soon can we expect delivery?

A1

Usually 3 to 5 working days.

Q2What is the condition of your machines?
A2In good working order.
Q3

What is your minimum quantity of the order?

A3

1 piece.

Q4

How can I get more details?

A4

We will be glad to help you out at WhatsApp/ Phone / Skype / WeChat


"Quality assurance" is our eternal pursuit and promise. Don't be hesitant to contact us! Welcome to place your orders of Mindray BeneHeart R3 ElectrocardiograPHiliphhilip ECG Machine patient monitor Parameter Board 115-018611-00. Any inquiry will be highly appreciated and thanks for your support.

China Mindray BeneHeart R3 ElectrocardiograPHiliph ECG Machine Parts Patient Monitor Parameter Board 115-018611-00 supplier

Mindray BeneHeart R3 ElectrocardiograPHiliph ECG Machine Parts Patient Monitor Parameter Board 115-018611-00

Inquiry Cart 0