►Brief Introduction
The M4735A is powered by a rechargeable Sealed Lead Acid (SLA)
battery (M3516A). Proper care of these batteries will ensure that
they have the energy required to operate the M4735A and deliver the
appropriate therapy. The M3516A battery differs from similar
batteries in that it is capable of providing much higher current
without causing an internal protection device to open and
disconnect the battery.
The defibrillator will take longer to charge when powered by AC
without a battery installed. To ensure optimal performance, always
have a fully charged battery in the defibrillator, even when using
AC power.
The repair PhilipilipilosoPhilipilipy for the SLA battery (M3516A)
is unit replacement. If a battery fails, it is replaced, not
repaired.
► Power Indicator Matrix
INDICATORS | SYSTEM STATE | MEANING | |||
AC POWER | BATTERY CHARGING | Connected to AC Mains? | Battery in Place? | Unit Power ON? | |
G = GREEN OR = ORANGE/AMBER | Y = YES N = NO --- = don’t care | ||||
OFF | OFF | N | --- | --- | Normal |
G | OFF | Y | N | --- | Normal |
G | OR | Y | Y | --- | Battery is charging. |
G | G | Y | Y | --- | Battery at least 90% charged. |
►Battery Capacity Test
The HeartStart XL uses the M3516A Battery Pack. It is rechargeable
sealed lead acid battery. Battery maintenance begins when you
receive a new battery and continues throughout the life of the
battery.
As sealed lead acid batteries age due to usage and storage
conditions, their capacity to store and deliver energy diminishes.
The Battery Capacity Test provides a quantitative measure of
battery capacity. Performing a Battery Capacity Test periodically
over the life of the battery gives you an indication of how that
battery is aging, and what level of performance it is capable of
delivering.
A battery’s capacity is most affected by the duration of use and
how deeply it is discharged before recharging, Therefore, some
general guidelines for performing the Battery Capacity Test are as
follow:
► To perform a Battery Capacity Test(CT).
Battery Capacity Test Results
If | Then |
Elapsed Time≥ 95 minutes and Low Battery time ≥ 10 minutes | 1. The battery passed the test 2. Record "Pass CT" and the date on the bottom of the battery 3. Recharge the battery before use. |
Elapsed Time< 95 minutes or Low Battery time < 10 minutes | 1. The battery failed the test. 2. Record "Fail CT" and the date on the bottom of the battery. 3. Discard the battery appropriately. |