Metal Shell 0.1μM Paint Thickness Tester Meter Gauge With Fe NFe Function

Brand Name:LEEB
Certification:CE, FCC EAC & ISO Certification
Model Number:Uee922
Minimum Order Quantity:1 Set
Delivery Time:In 7-10 work days
Payment Terms:T/T, PayPal, Western Union, L/C
Contact Now

Add to Cart

Verified Supplier
Location: Chongqing Chongqing China
Address: High&New Technology Industrial Park, Yubei District, Chongqing, China
Supplier`s last login times: within 1 hours
Product Details Company Profile
Product Details

Metal Shell 0.1μM Paint Thickness Tester Meter Gauge With Fe NFe Function


High Precision Metal Shell Coating Thickness Gauge With Fe And NFe Function


Functions & Features


● With high quality metal probes and shell.
● Large memory to save 4 types of materials and 1560 testing values.
● Software for PC connection and data transmission, analysis.
● Two measuring methods:continuous and single;
● Two working mode: direct and batch;
● Limit setting function.
● Switch off automatically or manually.
● 3 ways for easy Calibration: by one specimen, two specimens or five specimens to calibrate..
● Five statistics: Average, Maximum, Minimum, Testing times, Standard deviation.
● Standard delivery with 5 calibration specimens (48.5μm,99.8μm,249μm,513μm,1024μm)
● 3 years warranty and life-long services


Technical Parameters


Technical Parameters
Model No.Uee922
Measuring principleMagnetic induction (Fe ) & Eddy current ( NFe)
Measuring range (µm)0~1250μm
ProbeChangeable
ShellMetal
Accuracy±(2%H+1) μm; H refers to the thickness of testing piece
Minimum resolution (µm)0.1μm
Min curvature of the min area (mm)Convex1.5 Concave9
Diameter of the min area (mm)Φ7
Critical thickness of substrate (mm)0.5
Memory1560
Dimensions130*70*30mm
Power supply2*AA Alkaline battery
Standard Configuration
Main Machine, probe*1(Fe or NFe), substrate*1(Fe or NFe), software & USB, Calibration specimens*5, Users’ Manual, Qualified Certificate, Coin screwdriver, Packing list, Warranty card
Optional AccessoriesProbes,Specimens

Measuring Principle
This instrument adopts both magnetic and eddy current thickness methods
a) Magnetic method (F Probe)
Measuring Non-magnetic coatings on magnetic substrates (such as aluminum,chromium, copper, rubber coatings on steel, iron, alloy and hard magnetic steel substrates).
b) Eddy current method (N Probe)
Measuring Non-conductive coatings on Non-magnetic basis metals (such as enamel, plastic, rubber, paint coatings on copper, aluminum, zinc, tin substrates).

Calibration on Sandblast Surface
Sandblast the surface characteristics of led to the measured value deviates from the true value greatly, the coating thickness is roughly can use the following method to determine.
Method 1:
a) Instrument to use the method of 3.3.1 or 3.3.2.1 in radius of curvature and base material of the same smooth surface calibration.
b) Without plating in through the surface of the same treatment of measurement about 10 times, get the average Mo.
c) In the coating on the surface of measuring 10 times average Mm.
d) (Mm—Mo)±S is the coating thickness. S(standard deviation) is the bigger one between SMm and SMo.
Method 2:
a) Measured with the single mode under the directly way.
b) Use two specimens method to calibrate the instrument.
c) Measured 5 to 10 times on the standard piece.Press“STATS”,Statistical average value of the cladding thickness is.

PC Software: Software for PC connection and data transmission, analysis.



Packing box: High quality water-proof box



China Metal Shell 0.1μM Paint Thickness Tester  Meter Gauge With Fe NFe Function supplier

Metal Shell 0.1μM Paint Thickness Tester Meter Gauge With Fe NFe Function