VICTOR 11+ Process Signal Source handheld calibrator process calibrator multimeter digital calibrator

Brand Name:VICTOR
Certification:CE
Model Number:VICTOR 11+
Minimum Order Quantity:1pcs
Delivery Time:5 to 30 days
Payment Terms:T/T, Western Union, L/C, Paypal
Contact Now

Add to Cart

Site Member
Address: JINGYUAN 7 ROAD, JINGHE INDUSTRIAL PARK, NORTH DISTRICT,XI'AN,CHINA  710200
Supplier`s last login times: within 22 hours
Product Details Company Profile
Product Details

Process Signal Source LCD Thermocouple Simulator Calibrator VICTOR 11+


VICTOR 11+ Process Signal Source High Accuracy Output function Thermal resistance Thermocouple Frequency calibrator

  • Introduction

This process signals source (the calibrator in the following) is a handheld, battery-operated instrument that sources electrical and physical parameters.

Features:

l Source DC voltage, DC current, resistance, simulate Transmitter, TC, RTD, frequency, pulse and switch.

Others features:

  • Big LCD can display the TC/RTD measurement value and mV/Ω corresponding simultaneously; mA and mA% simultaneously.
  • TC source terminals and built-in lead connector of same temperature (RJ compensation with auto-reference joint point)
  • Manual step source and auto -step and sweeping –step source
  • Room temperature monitoring under any operation
  • Standard Accessories

Make sure that the package contains all the accessories listed below. And if you find they are damaged or any of them is missing, please contact the vendor from which you purchased the product as soon as possible. Refer to the replacing part list in 15.3 in the Manual if you want to order the replacing parts.

  • One set of Industrial testing Lead (CL727220)
  • A set of Testing Lead (Tp727110)
  • A set of Alligator clip (CC807130)
  • A quick reference guide
  • A User's Manual
  • One Fuse 50mA/250V

One Fuse 100mA/250V


Techincal Data


ProjectThe peference model
Performance
parameters

VICTOR 11+VICTOR 14+VICTOR 15+
SourceDC millivoltRange-10 ~ 1100mV-10.000~100.000mV-10 ~ 1100mV
Accuracy=
±(% of the set value+% of range)
0.02%+0.01%0.02%+0.01%0.02%+0.01%
Resolution1uV1uV0.01mV
DC voltageRange-1 ~ 11V-0.1 ~ 1V-1 ~ 11V
Accuracy=
±(% of the set value+% of range)
0.02%+0.01%0.02%+0.01%0.02%+0.01%
Resolution0.1mV10uV0.1mV
DC currentRange0 ~ 22mAN/A0 ~ 22mA
Accuracy=
±(% of the set value+% of range)
0.02%+0.02%0.02%+0.02%
Resolution1uA1uA
Sink currentRange0 ~ 22mAN/A0 ~ 22mA
Accuracy
±(% of the set value+% of range)
0.02%+0.02%0.02%+0.02%
Resolution1uA1uA
External power supply5 ~ 28V5 ~ 28V
ResistanceRange0 ~ 40KΩ0 ~ 4KΩN/A
Accuracy=
±(% of the set value+% of range)
0.02%+0.02%0.02%+0.02%
Resolution0.01Ω/0.1Ω/1Ω0.01Ω/0.1Ω
FrequencyRange1Hz ~100KHzN/A1Hz ~10KHz
Accuracy±2words±2words
Resolution0.01Hz/1Hz/0.1KHz/2KHz0.01Hz/1Hz/0.1KHz/2KHz
PulseFrequency1Hz ~10KHzN/A1Hz ~10KHz
Range1 ~100000cyc1 ~100000cyc
Accuracy±2words±2words
Resolution1cyc1cyc
SwitchRange1Hz ~100KHzN/AN/A
Accuracy±2words
Resolution0.01Hz/1Hz/0.1KHz/2KHz
ThermocoupleRange10types(R/S/K/E/J/T/B/N/L/U)10types(R/S/K/E/J/T/B/N/L/U)N/A
Accuracy0.4℃0.4℃
Resolution0.1℃/1℃0.1℃/1℃
RTDRange6types(Pt100/Pt200/Pt500/
Pt1000/Cu10/Cu50)
6types(Pt100/Pt200/Pt500/
Pt1000/Cu10/Cu50)
N/A
Accuracy0.2℃0.2℃
Resolution0.1℃0.1℃
MeasureDC millivoltRangeN/A-5 ~ 550mV-20 ~ 220mV
Accuracy=
±( % of Reading+ % of Range)
0.02%+0.01%0.02%+0.02%
Resolution1uV/10uV0.01mV
DC voltageRangeN/AN/A-5 ~ 55V
Accuracy=
±( % of Reading+ % of Range)
0.02%+0.01%
Resolution0.1mV/1mV
DC currentRangeN/AN/A-4 ~ 55mA
Accuracy=
±( % of Reading+ % of Range)
0.02%+0.01%
Resolution1uA
ResistanceRangeN/A0 ~ 5.5KΩN/A
Accuracy=
±( % of Reading+ % of Range)
0.05%+0.02%
Resolution0.01Ω/0.1Ω
FrequencyRangeN/AN/A3 ~50KHz
Accuracy±2words
Resolution0.01Hz/0.1Hz/1Hz
SwitchRangeN/AN/ACLOSE/OPEN
Threshold200 ~300Ω
PressureRangeN/AN/ADepends on the pressure module
Accuracy=
±( % of Reading+ % of Range)
Depends on the pressure module
Resolution5 digit display
Loop power supplyRangeN/AN/A24V
Accuracy±10%
The maximum current22mA
ThermocoupleRangeN/A10types(R/S/K/E/J/T/B/N/L/U)N/A
Accuracy0.7℃
Resolution0.1℃/1℃
RTDRangeN/A6types(Pt100/Pt200/Pt500/
Pt1000/Cu10/Cu50)
N/A
Accuracy0.5℃
Resolution0.1℃
Continuity beeperRangeN/A0 ~500Ω
Accuracy≤50Ω Sound≤250Ω Sound
Resolution0.01Ω
Other functionsBacklight
Auto Power-off
Charging function
Room temperature display
Battery detection
Communication functionRS232 serial communication, can realize remote control instrument by the device driver software
The mathematical treatmentData hold,zero,averageData hold,zero,averageData hold,zero,average
Characteristic
SafetyEN61010-1:2001EN61010-1:2001EN61010-1:2001
InsulationDC500V /50MΩDC500V /50MΩDC500V /50MΩ
EMCEN61326-1:2006EN61326-1:2006EN61326-1:2006
AuthenticationCECECE
DisplayDouble display on the LCD screenDouble display on the LCD screenDouble display on the LCD screen
Power supply4×1.5V AAA alkaline battery4×1.5V AAA Alkaline battery4×1.5V AAA Alkaline battery
Size205 × 95 × 42 mm205 × 95 × 42 mm205 × 95 × 42 mm
WeightAbout 500gAbout 500gAbout 500g
Standard Quantity Per Carton10pcs10pcs10pcs
Standard Carton Measurement395mm*385mm*260mm395mm*385mm*260mm395mm*385mm*260mm
Standard Carton Cross Weight10.4Kg10.3Kg11.5Kg

Product Details:


Packing And Accessories:

China VICTOR 11+ Process Signal Source  handheld calibrator process calibrator multimeter digital calibrator supplier

VICTOR 11+ Process Signal Source handheld calibrator process calibrator multimeter digital calibrator

Inquiry Cart 0