LCD Display Portable Reference Standard Meter

Brand Name:HNMC
Model Number:MCSB01
Minimum Order Quantity:1 pc
Delivery Time:20-25 Days
Payment Terms:T/T, L/C, D/P
Place of Origin:China
Contact Now

Add to Cart

Active Member
Location: Zhengzhou Henan China
Address: No.1207-1, Building #1, the National University Sci-Tech Park of Henan Province, No.11 Changchun Road, Zhengzhou, Henan 450000
Supplier`s last login times: within 38 hours
Product Details Company Profile
Product Details

0.02 Class Three Phase Standard Meter MCSB01


Technical Index


ParametersMeasuring rangeResolutionAccuracy
Voltage (P-N)3*(1V ~ 576V) (auto & manual mode)6 valid digits0.05% OR 0.02% (40-576V)
Current3 x (1mA ~ 120A), auto & manual mode6 valid digits0.05% or 0.02% (10mA-120A)
Active Power / Energy40V ~ 576V / 10mA ~ 120A / PF ≥ 0.56 valid digits0.05% or 0.02%
Reactive & apparent Power / Energy40V ~ 576V / 10mA ~ 120A / PF ≥ 0.56 valid digits0.05% or 0.02%
Power factor (PF)-1.0 ~ +1.06 valid digits±0.0005
Phase angle0 º ~ 360º6 valid digits0.02º (40-576V,10Ma-120A, sine wave)
Frequency45Hz ~ 65Hz6 valid digits0.005%

Work Condition


ItemSpecification
Temperature10ºC~40ºC (calibration temperature 23±1ºC)
Humidity45~80%R.H
Power supply220V±10%, 50/60HZ
Power consumption≤50VA
Weightapprox 7.5kgs

Features

1) Small weight & size, easy to use in lab and at work site.

2) LCD display

3) Can work as standard meter and as calibrator of meter test equipment & bench

4) Can do calibration & test of energy meter test bench & equipment, and can cover all the

necessary test items except withstand voltage test

5) Adopting DSP technology and temperature auto balance and other compensation technology,

it ensure a stable parameters and performance

6) Can do below various test:

- Four-quadrant test of AC voltage, current, power (active, reactive, apparent) and energies

under various wiring ways including 1P2W, 3P3W, 3P4W and etc.

- Reactive power and energy test under real reactive, cross-phase & neutral point reactive, etc

- Test of voltage, current, phase, frequency, active, reactive, power factor and error, and etc.

- Test of voltage, current and power stability (5 kinds of calculation method of power stability)

- Test of three phase amplitude asymmetry of voltage and current

- Test of three phase unbalance of voltage and current

- Test of phase asymmetry

- Space magnetic induction intensity test

- Dotted terminal voltage dip test

- Constant & frequency test of meter under test

- Basic wave power test

- Full power test (basic wave power and harmonic wave power of each time)

- Demand test

7) Can test error of three kinds of different type meters at the same time, including active test,

reactive test, apparent test, voltage and current test

8) Also can test comprehensive error of equipment comprising of three pieces single phase meter

9) Also can work as three pcs single phase standard meter

10) With three impulse outputs

11) Constant can be set separately (auto mode or manual mode) and type of constant can be

set separately, such as active, reactive, apparent, voltage and current.

12) Can calibrate the energy error calculation and standard deviation error calculation of meter

test equipment & bench and energy meter; and constant of meter under test & pulse quantity can

be set automatically.

13) Vector diagram display with various ways

14) Can display waveform, spectrum (with zoom function), amplitude & content & phase of harmonic wave

15) Can directly display working status and various test data (voltage, current, frequency,

power, phase angle, error,etc)

16) Can display waveform, phase vector diagram and harmonic wave diagram of test voltage & current

17) Harmonic analysis and distortion calculation

18) With store functions, can save at least 30 groups test records.

19) Communication port: RS232, USB. Baud rate can be set when adopting asynchronous communication.



China LCD Display Portable Reference Standard Meter supplier

LCD Display Portable Reference Standard Meter

Inquiry Cart 0