Product Details
[Main Ingredients]cTnI.pdf
Name | Loading capacity | Ingredient |
Test card | 25 | The product consists of fluorescent mat (coated with
fluorescently-labeled cTnI monoclonal murine antibody),
nitrocellulose membrane (coated with cTnI monoclonal murine
antibody and goat anti mouse IgG antibody), absorbent paper and
bottom lining. |
Sample diluent | 25 | Phosphate buffer |
ID card | 1 | Record standard curve information of this batch of reagents |
[Sample Requirements]
- Serum and EDTA·Na2 anticoagulant plasma and whole blood, EDTA·K2 anticoagulant plasma and whole blood, sodium citrate anticoagulant
plasma and whole blood can be used.
- Collect venous blood using conventional laboratory method and avoid
hemolysis in the treatment process.
- Clinical samples should be tested at room temperature (15-30)℃ in
4h after collected. Whole blood specimens can be stored at (2-8)℃
for 24h and should not be frozen; serum or plasma specimens can be
stored at (2-8)℃ for 7 days and at -20ºC for 30 days.
- The sample must be re-warmed to room temperature (15-30)℃ before
test. Frozen samples should be completely melted, re-warmed and
mixed before use and should not be frozen repeatedly.
- Please do not test samples of severe hemolysis, severe lipoidemia
and icterus.
[Test Method]
- Please thoroughly read the specification before test. Frozen test
card and sample should be placed at room temperature (15-30)℃ for
at least 30min before use.
- Start NIR-1000 dry fluoroimmunoassay analyser and verify quality
control according to the specification. (Note: Reagent has been
calibrated in advance and calibration curve parameters of each
batch of reagents have been stored in the information card. Insert
the information card before use and carry out test without
re-calibration after passing quality inspection; otherwise,
identify the cause before test.)
- Take out the test card from the aluminum foil bag and use it within
15min
- Place the test card on a clean horizontal table top and label it.
- Serum, plasma or whole blood specimen: Take 100μL of sample and add
it into 300μL of buffer solution (1:3). Then, mix the solution
evenly, take 100μL of the solution and add it into the test card
well.
- Insert the test card into NIR-1000 dry fluoroimmunoassay analyser
and press “Timing test” to keep time for 12min automatically. The
analyser will judge and read the test result automatically and
display it in the screen. Or insert the test card into the analyser
after 12min and press “Instant test”, the instrument will judge and
read the result automatically.