[Product Name]cTnI.pdf
Cardiac Troponin I (cTnI) Rapid Quantitative Test(Fluorescence
immunoassay)
[Intended Use]
The product is used to determine the content of cardiac troponin I
(cTnI) in whole blood, plasma and serum of human body and is mainly
used clinically for auxiliary diagnosis of myocardial infarction.
[Test Principle]
The kit uses immunochromatographic assay. First, cTnI antigen in
the sample combines with the fluorescently-labeled cTnI monoclonal
antibody conjugate. Then, it continues to move and combines with
another cTnI monoclonal antibody fixed on the nitrocellulose
membrane to form double-antibody sandwich immune complex in the
position of the nitrocellulose membrane test line and analyze and
obtain quantitative test result using NIR-1000 dry
fluoroimmunoassay analyser.
[Precautions]
- Test card and buffer solution are single-use and they cannot be
reused.
- Please inspect packaging integrity and validity of kit before use
and then unpack the product. If the product is stored at low
temperature, restore to room temperature (15℃-30℃) before unpacking
and use. Reagent cannot be used if packaging is damaged and the
validity period expires.
- Take the test card out of the aluminum foil bag and carry out
experiment in 15min. Do not place it in the air for a long time to
avoid dampness.
- It is required to strictly comply with the requirements for sample
collection and storage. If the sample is turbid, please centrifuge
and precipitate it before use.
- The kit contains products from animals. Eligible information about
animal source and sanitary condition cannot absolutely ensure
inexistence of infectious pathogen. Therefore, these products
should be disposed of as latent infective material, and all
samples, reagents and latent contaminants should be disinfected and
disposed of according to relevant local regulations.
- Too high or too low hematocrit of red cells may affect whole blood
test result, so verification should be conducted using other
methods .