TD1050A Calibration Device for Clamp Meter

Introduction

  • TD1050A, a calibration device for clamp ammeters. The optional accuracy are 0.02% and 0.05%.

  • The device calibrates the AC/DC clamp ammeters by single coil with high accuracy and easy traceability, and has the functions of AC/DC current output, resistance simulation, frequency/phase adjustment, power detection.

  • Meet the requirements of JJF 1075-2015 Calibration Specification for Clamp Ammeter. (Tunkia is involved in drafting)

Features

  • Calibrate AC/DC clamp ammeters by single coil: The AC/DC current source connect to a single coil semi-circular copper ring with a diameter of 1m, can directly output AC/DC high current up to 2200 A (or 1100 A). Suitable for calibrating AC/DC clamp ammeters with high accuracy.

  • Place the clamp meter in the center of the copper ring when calibrating. The diameter of the copper bar in the center position is smaller than the other positions, which is convenient for the access of small size clamp meters.

  • Easy to open the copper ring, and thread a standard current transformer or I-V converter into the ring for calibrating the device itself regularly.

  • Reserved terminals for high current, to provide current for the calibration of DC shunts.

  • Value Adjustment Method: Fixed-point, rotary encoder, and step.

  • Frequency and Phase (Option): The AC output frequency and the phase between voltage and current is variable.

  • Power Output (Option): Test DC power, active power, reactive power, apparent power and power factor.

  • Proprietary Software (Option): Support the calibration of the checked ammeters by semi-automatic or full-automatic, data management and report export.

Application

Calibrated Types

Accuracy of Calibrated Device

0.02% Calibrator

0.05% Calibrator

AC/DC Ammeters

0.1 and below

0.2 and below

AC/DC Voltmeters

0.1 and below

0.2 and below

DC Resistance

0.2 and below

0.2 and below

AC/DC Power

0.2 and below

0.5 and below

DC Shunts

0.1 and below

0.2 and below


Specification


AC Voltage Output

RangeStability (%/min)

Accuracy (k = 2)

(ppm*RD + ppm*RG)①

Max. Load

Current (mA)

0.02%0.05%0.02%0.05%
10 V0.0050.01120 + 80300 + 200800
30 V0.0050.01120 + 80300 + 200700
100 V0.0050.01120 + 80300 + 200200
300 V0.0050.01120 + 80300 + 20070
750 V0.0050.01120 + 80300 + 20030
Remarks: ①: RD is the reading,RG is the range
  • Output Range of Voltage: 1 V ~ 825 V

  • Display: 6 digits

  • Resolution: 0.002 %*RG

  • Distortion: <0.5 %

  • Protection: Short-circuit protection, overload protection, overheating protection.


AC Current Output

ModelRangeResolutionStability (%/min)

Accuracy (k = 2)

(ppm*RD + ppm*RG)

Max. Load

Voltage (V)

0.02%0.05%0.02%0.05%
Small current loop100 mA1 μA0.010.01120 + 80300 + 20010
300 mA1 μA0.010.01120 + 80300 + 20010
1 A10 μA0.010.01120 + 80300 + 20010
3 A10 μA0.010.01120 + 80300 + 20010
10 A100 μA0.0050.01120 + 80300 + 20010
30 A100 μA0.0050.01120 + 80300 + 2005
100 A1 mA0.0050.01120 + 80300 + 2005
High current loop250 A1 mA0.010.01120 + 80300 + 2007
500 A1 mA0.010.01120 + 80300 + 2007
1000 A10 mA0.0050.01120 + 80300 + 2007
2000 A②10 mA0.0050.01120 + 80300 + 2007
Remarks: ②: 2100 A for option
  • Output Range of Current: 10 mA ~ 1100 A (2100 A);

  • Range Switching: Automatic or Manual

  • Resolution: 0.002 %*RG

  • Distortion: <0.5 % (except 1 kA/2 kA); <1 % @ 1 kA/2 kA range

  • Protection: Short-circuit protection, overload protection, overheating protection.


Frequency

TypeRangeAdjust finenessAccuracy (k = 2)
Frequency

40.000 Hz ~ 400.000 Hz

(1 kHz can be customized)

0.001 Hz0.005%*RD

DC Voltage Output

RangeStability (%/min)

Accuracy (k = 2)

(ppm*RD + ppm*RG)

Max. Load

Current (mA)

0.02%0.05%0.02%0.05%
200 mV0.0050.01120 + 80300 + 200100
1 V0.0050.01120 + 80300 + 200100
10 V0.0050.01120 + 80300 + 200100
30 V0.0050.01120 + 80300 + 200600
100 V0.0050.01120 + 80300 + 200200
300 V0.0050.01120 + 80300 + 20060
600 V0.0050.01120 + 80300 + 20025
1000 V0.0050.01120 + 80300 + 20020
  • Output Range of Voltage: 20 mV to 1100 V

  • Resolution: 0.002 % of range

  • Ripple Coefficient: <1 %

  • Protection: Short-circuit protection, overload protection, overheating protection.


DC Current Output

ModelRangeResolutionStability (%/min)

Accuracy (k = 2)

(ppm*RD + ppm*RG)

Max. Load

Voltage (V)

0.02%0.05%0.02%0.05%
Small current loop100 mA1 μA0.0050.01150 + 50300 + 10010
300 mA1 μA0.0050.01150 + 50300 + 10010
1 A10 μA0.0050.01150 + 50300 + 10010
3 A10 μA0.0050.01150 + 50300 + 10010
10 A100 μA0.0050.01150 + 50300 + 10010
30 A100 μA0.0050.01150 + 50300 + 1005
100 A1 mA0.0050.01150 + 50300 + 1005
High current loop250 A1 mA0.0050.01150 + 50300 + 1005
500 A1 mA0.0050.01150 + 50300 + 1005
1000 A10 mA0.0050.01150 + 50300 + 1005
2000 A②10 mA0.0050.01150 + 50300 + 1005
Remarks: ②: 2100 A for option
  • Output Range of Voltage: 20 mA ~ 1100 A (2100 A);

  • Resolution: 0.002 %*RG

  • Range Switching: Automatic or Manual

  • Distortion: <0.5 % (except 1 kA/2 kA); <1 % @ 1 kA/2 kA range

  • Protection: Short-circuit protection, overload protection, overheating protection.


Power Output (Option)

TypeRangeAccuracy (k = 2), (%*FS)③
0.02%0.05%
DC PowerCombination of DC voltage and current range.0.050.1
Active Power |cosφ|≥0.5Combination of AC voltage and current range.0.050.1
0.50.050.1
Reactive Power |sinφ|≥0.50.050.1
Power Factor-1.0000…0.0000…+1.00000.050.1
0.050.1
Remark: ③: FS = Voltage range × Current range

Resistance

RangeResolutionAllowable Current

Accuracy (k = 2)

(ppm*RD + ppm*RG)

100 Ω1 mΩ300 + 200300 + 200
1 kΩ10 mΩ300 + 200300 + 200
10 kΩ100 mΩ300 + 200300 + 200
100 kΩ1 Ω300 + 200300 + 200
1 MΩ10 Ω300 + 200300 + 200
10 MΩ100 Ω600 + 400600 + 400
100 MΩ1 kΩ10 nA ~ 50 nA3000 + 2000
  • Output Range of Resistance: 10 Ω ~ 110 MΩ

  • Display: 6 digits