>>Technical Parameter
Type: PVC membrane ion-selective electrode
Measuring range: 0.4-62000 mg/L
pH range: 2.5-11 pH
Temperature range: 0-50℃
Pressure resistance: Non-pressure resistant
Temperature sensor: NTC10K/NTC2.2K/PT100/PT1000
Housing material: PP
Connection thread: NPT3/4’’
Cable length: 10m or as agreed
Cable terminal: Tinned or as agreed
>>Overview
The ion-selective electrode (ISE), also known as a membrane
electrode, is an electrode with a sensitive membrane that generates
a membrane potential based on ion exchange or diffusion. It
primarily consists of an ion-selective membrane, an internal
reference electrode (typically Ag/AgCl), and an internal reference
solution (composed of chloride and strong electrolyte solutions of
the target ion). The ISE measures the electrical potential of a
specific ion in a solution, which is determined relative to a
reference electrode with a constant potential. The potential
difference between the two electrodes depends on the activity of
the specific ion in the solution. This activity is related to the
ion’s concentration, enabling quantitative analysis.
>>Feature
- Reagent-free, pollution-free, cost-effective, and eco-friendly.
- Redox potential measurement method ensures fast response time and
stable signals.
- Electrode made of PP (polypropylene) material.
- High-quality sensor-specific cables enhance signal accuracy and
stability.
>>Pre-Measurement Preparation
- Remove the protective cap from the electrode tip. Note: Avoid touching the sensitive area with fingers.
- Shake the electrode downward (like a thermometer) to remove
internal air bubbles.
- For new or long-unused electrodes, perform activation by soaking in
1 ppm standard solution for at least 24 hours.
- After activation, rinse the electrode with deionized water and
gently blot dry with a non-woven cloth or clean tissue. Before
testing, immerse the electrode tip in deionized water for 5 minutes with stirring. Replace the water multiple times and repeat
rinsing to prevent measurement errors.
>>Calibration (Slope Determination)
Connect the electrode to the meter.
Pour 100 mL of deionized water into a 150 mL beaker and stir thoroughly. Set the meter to mV mode.
Blot the rinsed electrode dry and immerse it in the prepared
deionized water.
Add 1 mL of 1000 ppm standard solution to the beaker, stir thoroughly, and record the stable reading (E₁).
Add 10 mL of the same standard solution to the beaker, stir thoroughly, and record the stable reading (E₂).
The slope at 25°C is calculated as E₂ – E₁. Compare the result with the specifications to validate electrode
performance.
>>Troubleshooting
If the slope deviates from the specified range:
- Prepare freshly diluted standard solution (lowest concentration).
- Soak the electrode in the diluted solution for 10 minutes before testing.
- Repeat the calibration steps.
If the issue persists, contact the manufacturer’s technical support
or replace the electrode.
>>Storage
- Short-term storage: Preserve the electrode in deionized water or low-concentration
standard solution (1 ppm). No reactivation is required before
reuse.
- Long-term storage: Rinse the electrode with deionized water, gently wipe the
membrane with a non-woven cloth, dry thoroughly, and store in its
original packaging. Reactivate in 1 mg/L standard solution, deionized water, or low-concentration
solution for over 2 hours before reuse.
Company Profile
Founded in 2015, Juge focuses on the field of environmental online
monitoring and has gathered outstanding talents who have been
engaged in environmental online monitoring for many years. At the
same time, it cooperates with some key professional universities
and research institutes in China, and is a national high-tech
enterprise that integrates research and development, production,
sales, and after-sales service with environmental online monitoring
systems as its core. It is a provider of overall solutions for
smart environmental online monitoring systems.
Main business focuse on water quality monitoring instruments, flue
gas analyzers, VOC online monitoring systems, CEMS flue gas
continuous monitoring systems, and flue gas particulate matter
monitoring systems. Widely used in safety, environmental
protection, petroleum, chemical, pharmaceutical, metallurgical,
coal chemical, warehousing and other fields
With a pragmatic enterprise management team, high-end R&D
talents, experienced manufacturing experts, and enterprising
marketing elites, we have developed a leading position among
domestic peers and have always adhered to the business philosophy
of providing users with higher quality products and more intimate
services. We have made a vision of being responsible and doing our
best to create an environmentally friendly and healthy work and
living environment, and our goal is to pursue the interests of our
customers.