0-50mg/L Online Phosphate Analyzer Water Analyzer Instrument RS485 interface

Brand Name:JUGE
Certification:IOS9001, CE
Model Number:AG9282C
Minimum Order Quantity:1 Unit
Delivery Time:5-8 working days
Payment Terms:Online order, T/T
Contact Now

Add to Cart

Verified Supplier
Location: Nanjing China
Address: 2F, BLK A, TAIPING IND PARK, #219 JIANGJUN AVE, JIANGNING DISTRICT, NANJING, CHINA
Supplier`s last login times: within 39 hours
Product Details Company Profile
Product Details

Online Water Quality Phosphate Analyzer


>>Application
With the increasing standards of power plants and the requirements of desalination and water purification projects, the timeliness and accuracy of phosphate monitoring have significantly improved. Particularly, the widespread adoption of low-phosphorus treatment technology has made the measurement accuracy of low-concentration phosphate even more critical. The AG9282C Phosphate Analyzer developed by our company utilizes a photometric dosing method, cold light source spectrophotometric measurement, and a dead-volume-free reaction cell, ensuring precise and reliable instrument performance.


>>Features
▪ ‌Photometric dosing method‌ enables precise reagent measurement.
▪ ‌Cold light source spectrophotometry‌ extends the lifespan of the light source.
▪ ‌Automatic light intensity adjustment‌ ensures measurement accuracy even after light source attenuation.
▪ ‌Large-capacity memory‌ stores measurement data for up to 5 years.
▪ ‌7-inch touchscreen color LCD‌ offers intuitive operation and display.
▪ ‌4 isolated current outputs‌ configurable for any channel, range, or PID.
▪ ‌4 relay outputs‌ programmable for over-limit alarms, sample-loss alarms, or system fault alarms.
▪ ‌RS485 interface‌ supports remote data monitoring.
▪ ‌Historical data and trend curves‌ accessible for any specified time period.


>>Technical Parameter

Specification

Detail

Measurement range

0-50mg / L OEM

Instrument display value error

±1%F.S

Repeatability

≤1%

Measurement period

minimum of 20 min

Stability

Baseline drift:Use a blank calibration without compromising accuracy

Chemical drift:not more than 1% reading / month (depending on reagent stability)

Type of reagents

1 kinds

Temperature Sensor

PT 100

Temperature measurement range

(0.0~60.0) ℃

Temperature measurement accuracy

± 0.5℃

Temperature compensation range

(0.1~60.0) ℃

Isolate the output signal and load

4 ~ 20 mA (RL <500 Ω)

Alarm contact capacity

normally open contact AC220V / 3A, DC24V / 5A

Working conditions

Temperature:0~45℃

Humidity: ≤85%

Sample conditions

Flow rate:100-300 mL/min

Temperature:(5~50)℃

Pressure:14 KPa

Permissible solid composition of water sample

≤ 5 microns (no glue is allowed, no grease)

Power supply

AC220V± 10% 50Hz±0.5Hz power: 150W

Instrument external dimensions

height, 720, width, 460, depth, 360 (mm)

Instrument opening size

665x410mm


>>Dimension

China 0-50mg/L Online Phosphate Analyzer Water Analyzer Instrument RS485 interface supplier

0-50mg/L Online Phosphate Analyzer Water Analyzer Instrument RS485 interface

Inquiry Cart 0