Online Water Quality Hydrazine Analyzer


>>Application
To ensure the deaeration efficiency of the water supply system, hydrazine is added after the deaerator to achieve chemical deaeration. This requires maintaining a surplus of hydrazine in the system, which imposes higher demands on the timeliness and accuracy of hydrazine monitoring. The AG9267C Hydrazine Analyzer developed by our company adopts a photometric dosing method, cold light source spectrophotometry, and a dead-volume-free reaction cell, ensuring precise and reliable measurement.


>>Features

  • Photometric dosing method‌ enables accurate chemical measurement.
  • Cold light source spectrophotometry‌ prolongs the service life of the light source.
  • Automatic light intensity adjustment‌ ensures measurement accuracy even as the light source ages.
  • Temperature-controlled reaction system‌ guarantees stable measurement and calibration.
  • Large-capacity memory‌ stores measurement data for up to 5 years.
  • 7-inch touch color LCD screen‌ provides intuitive operation and clear display.
  • 1 isolated current output‌ configurable for any channel, range, or PID control.
  • 1 relay output‌ configurable for over-limit alarms, sample loss alarms, or system fault alerts.
  • RS485 interface‌ supports remote data monitoring.
  • Historical data query‌ allows retrieval of measurement curves and alarms for any time period.

>>Technical Parameter

Specification

Detail

Measurement range

0-100 µg / L

Instrument display value error

±2%F.S

Repeatability

≤1%

Measurement period

minimum of 25 min

Stability

Baseline drift:Use a blank calibration without compromising accuracy

Chemical drift:not more than 1% reading / month (depending on reagent stability)

Type of reagents

1 kind

Temperature Sensor

PT 100

Temperature measurement range

(0.0~60.0) ℃

Temperature measurement accuracy

± 0.5℃

Temperature compensation range

(0.1~60.0) ℃

Isolate the output signal and load

4 ~ 20 mA (RL <500 Ω)

Alarm contact capacity

normally open contact AC220V / 3A, DC24V / 5A

Working conditions

Temperature:0~45℃

Humidity: ≤85%

Sample conditions

Flow rate:100-300 mL/min

Temperature:(5~50)℃

Pressure:14 KPa

Permissible solid composition of water sample

≤ 5 microns (no glue is allowed, no grease)

Power supply

AC220V± 10% 50Hz±0.5Hz power: 150W

Instrument external dimensions

height, 720, width, 460, depth, 360 (mm)