BGD 256 Digital Single-Beam UV-Vis Spectrophotometer
1.Description
The principle of the spectrophotometer is a light source generated
by a continuous radiation light by the dispersion system
(monochromator) decomposed into wavelengths of continuous
monochromatic light, monochromatic light by irradiation to the
sample cell, through the pool of samples will be absorbed by the
detection of the intensity of the incident light S and the
intensity of the incident light R, the transmittance T = S / R. If
the wavelength is continuously adjusted (i.e., scanning), it can be
obtained in the biochemical samples on the wavelength distribution
of the transmittance, its transmittance on the wavelength
distribution and the composition of the sample and concentration
related. The transmittance of a biochemical sample is distributed
about the wavelength, and the distribution of the transmittance
about the wavelength is related to the composition and
concentration of the sample. Thus, the composition and
concentration of the sample can be determined.
Spectrophotometers are commonly used for substance identification,
purity checking, and the study of organic molecular structure. In
the quantitative aspect, it can determine the content of each
component in compounds and mixtures with complex structure, and can
also determine the dissociation constant of substances, the
stability constant of complexes, the molecular weight
identification of substances and the indication of endpoints in
microtitration, and be used as a detector in high-performance
liquid chromatography, etc. The spectrophotometer can be used to
determine the concentration of a sample.
1200L/mm UV grating, C-T monochromator structure, sample chamber
can be extended to 10cm sample holder, automatic wavelength
setting, automatic adjustment of 0%T/100%T, T/A conversion. With
automatic correction of wavelength error, light source automatic
switching and other functions; with photometric, quantitative and
other functions; with RS232 communication interface can be optional
printer.
2.Technical Parameter
Wavelength Drive | Automatic |
Wavelength Range | 190 nm - 1100 nm |
Wavelength Accuracy | ±1 nm |
Wavelength Repeatability | 0.5 nm |
Spectral Bandwidth | 4 nm |
Transmittance Range | 0-200% T |
Transmittance Accuracy | ≤0.3% T |
Transmittance Repeatability | 0.1% T |
Absorbance Range | -0.4-4.0 A |
Concentration Display | 0-99,999 |
Stray Light | ≤0.1% T |
Stability | ±0.001 A/hour |
Dimensions (L×W×H) | 620 mm × 620 mm × 380 mm |
Net Weight | 21 kg |