4C2H5OH-100 4-Series Electrochemical Gas Sensor Gas Leak Monitoring And Alarm
Declaration:
1. 4-Series electrochemical gas sensors are designed for gas leak
monitoring and alarm. They are not
suitable for long-term exposure to the gas to be measured or
long-term on-line continuous detection
and analysis instruments.
2. Sensor pins must be connected through a PCB socket, Welding the
pin directly will damage the
sensor.
3. Sensors should avoid contact with organic solvents, alcohol,
paint, oil and high concentration of
gas, including silica gel and other adhesives.
4. Electrochemical sensors with positive output current (such as
CO, H2S, SO2, NH3, etc.) require
oxygen to participate in the reaction and should be calibrated with
the standard gas of air as the
background gas and test, otherwise it will destroy the sensor's
performance.
5.Do not remove or insert the sensor when it is powered on.
Otherwise, the sensor may be damaged
or abnormal.
Repeatability | 1% output signal |
Linearity | Regression coefficient R2 = 0.999 |
Operating Temperature Range | -20℃ ~ +50℃ |
Operating Humidity Range | 15% ~ 90%RH (relative humidity), non-condensing |
Operating Pressure Range | Atmospheric pressure ±10% |
Pressure Factor | < 0.02% signal / MBar |
Baseline Extent | < ±0.4 μA |
Long-Term Drift | < 5% signal loss / year |
Bias Voltage | +300 mV (Note: The biased sensor must be calibrated after more than 12 hours of stability after being inserted into the instrument) |
Load Resistance | 10 Ω |
Built-In Filters | No |
Lifetime (in Air) | 2 years |
Warranty Period | 1 year (from the date of shipment) |
Output Current Polarity | Positive current (define the output current of carbon monoxide and hydrogen sulfide sensor as positive) |