SGX-4OX Electrochemical Gas Sensor Gas Oxygen Sensor 0-100% Oxygen (O2) Sensor With On Board Temperature Compensation For Ventilators

Model Number:SGX-4OX
Minimum Order Quantity:1
Delivery Time:3-5work days
Payment Terms:L/C, D/A, D/P, T/T, Western Union, MoneyGram
Output signal:9-13mV in Air*
Zero current (offset):<0.1% O2
Contact Now

Add to Cart

Verified Supplier
Location: Shenzhen Guangdong China
Address: Room 3306AB 33rd Floor, SEG Plaza,Huaqiang North Road, Futian District, Shenzhen Guangdong Province 518028
Supplier`s last login times: within 26 hours
Product Details Company Profile
Product Details

SGX-4OX Electrochemical Gas Sensor Gas Oxygen Sensor 0-100% Oxygen (O2) Sensor With On Board Temperature Compensation For Ventilators


CROSS - SENSITIVITY DATA:

Toxic gases at TLV levels will have no cross-sensitivity effect on SGX-VOX oxygen sensors. At very high levels (i.e. % levels), highly oxidising gases (e.g. ozone, chlorine) will interfere to the extent of their oxygen equivalent, but most other commonly occurring gases will have no effect.


ACID GASES :

Acid gases such as CO2 and SO2 will be absorbed by the electrolyte and tend to increase the flux of oxygen to the electrode. This gives an enhanced oxygen signal of approximately 0.3% of signal per 1% CO2. SGX-VOX sensor is not suitable for continuous operation in concentrations of CO2 above 25%. The SGX-VOX sensor is not designed for use in applications where anaesthetic gases are present.


Temperature Range-20°C to +50°C
Pressure Range500 to 2000 mbar
Operating Humidity Range0% to 99% RH (non condensing)
Long Term Output Drift<5% per annum
Recommended Storage Temp0°C to 20°C


China SGX-4OX Electrochemical Gas Sensor Gas Oxygen Sensor 0-100% Oxygen (O2) Sensor With On Board Temperature  Compensation For Ventilators supplier

SGX-4OX Electrochemical Gas Sensor Gas Oxygen Sensor 0-100% Oxygen (O2) Sensor With On Board Temperature Compensation For Ventilators