CNLH Oxygen (O2) Gas Sensor Part Numbers: AA015-000 For Oxygen Analyzers 2-Electrode Electrochemical
Document Purpose:
The purpose of this document is to present the performance
specification of the CNLH oxygen gas sensor.
This document should be used in conjunction with the Operating
Principles (OP02) and the Product Safety Datasheet
(PSDS 4).
The data provided in this document are valid at 20°C, 50% RH and
1013 mBar for 3 months from the date of sensor
manufacture.
Output signal can drift below the lower limit over time. For
guidance on the safe use of the sensor, please refer to
the Operating Principles (OP02).
* Specifications are valid at 20°C, 50% RH and 1013 mBar, using
City Technology recommended
circuitry. Performance characteristics outline the performance of
sensors supplied within the
first 3 months. Output signal can drift below the lower limit over
time.
Operating Temperature Range | -20°C to +50°C |
Temperature Coefficient | 0.2% signal/°C |
Recommended Storage Temp | 0°C to 20°C |
Operating Pressure Range | Atmospheric ± 10% |
Operating Humidity Range | 0 - 99% RH non-condensing |
Long Term Output Drift | <5% signal loss/year |
Expected Operating Life | See below |
Storage Life | 6 months in CTL container |