3HYE Hydrogen 3 series electrochemical gas sensor 0-10000ppm
The output of a CiTiceL can vary with
temperature. The graph here shows the variation
in output with temperature for 3HYE CiTiceLs
based on a sample of about 16 sensors. The
results are shown in the graph as a mean for the
batch, and expressed as a percentage of the
signal at 20°C.
From a statistical viewpoint, for a sample of
this size, the range in values observed for all
sensors of this type will fall within a range three
times the standard deviation above or below the
mean. Assuming therefore this sample is typical,
then the temperature behaviour of all 3HYE
CiTiceLs will fall in the band +3SD to -3SD.
Maximum Zero Shift (+20°C to +40°C) | -150ppm equivalent |
Long Term Output Drift | <2% signal loss/month |
Bias Voltage | Not required |
Repeatability | 2% of signal |
Weight | 22g |